The chemicals contained in this kit may be hazardous if

improperly handled. Read the relevant Health and Safety

Data Sheet before performing this test.

HEALTH AND SAFETY

HEALTH AND SAFETY

HEALTH AND SAFETY

HEALTH AND SAFETY

HEALTH AND SAFETY

R E F E R E N C E S

R E F E R E N C E S

R E F E R E N C E S

R E F E R E N C E S

R E F E R E N C E S

P.Sequi,

Chimica del suolo

, Patron editore, Ed. 1991

Interferences

• Compounds with buffering capacity (substances that

keep the pH of sample at a constant value, e.g.

phosphate buffer, ammonia buffer): if after Mannitol

reagent has been added pH of sample does not change,

verify if pH increases a little (0.05 pH units) by adding

HI 38074E-0 reagent drop by drop. In this case boron

may be present, but a buffer interferes with boron

determination. It is possible to have only an indication

for boron concentration. Use the Boron Conversion Table

below to convert the number of drops added in mg/L

(or ppm) of Boron and record the concentration of boron

in your sample as lower than the value found (e.g.: 2

drops correspond to 0.4 mg/L of Boron, thus boron

concentration in your sample is lower than 0.4 mg/L).

• Bicarbonates: if hardness of sample is about 150 ppm

CaCO

3

, bicarbonates are present at pH 8.3, and they

interfere with boron determination. In this case the

sample needs to be pretreated before analysis as de-

scribed in the following paragraph.

Pretreatment to eliminate carbonate interference

• Fill the plastic vessel with 50 mL of sample and pour it

into the empty bottle.

• Add HI 38074B-0 reagent until pH is lower than 4.5.

Place the cap, but DO NOT SCREW IT.

• Heat the sample for about 45 seconds in a microwave

oven at 500W (do not boil the sample). Swirl several

times to eliminate air bubbles and allow the sample to

cool down to room temperature. Pour the sample into

the plastic vessel and perform the test as described

above.