23
mV CALIBRATION
(for technical personnel only)
All the instruments are factory calibrated for mV.
Hannas ORP electrodes are interchangeable and no mV calibration is
needed when they are replaced.
If the mV measurements are inaccurate, mV recalibration should be performed.
For an accurate recalibration, contact your dealer or the nearest Hanna
Customer Service Center, or follow the instructions below.
A two or three-point calibration can be performed at 0.0 mV, 600.0 mV and
1800.0 mV.
Attach to the BNC connector a mV simulator with an accuracy of ±0.1 mV.
With the instrument off, press and hold down the
CFM
&
RCL
keys, then
power on the instrument. The CAL tag will appear and the secondary
LCD will show 0.0 mV.
Set 0.0 mV on the simulator.
When the reading is stable and close to the selected calibration point,
READY tag will appear and CFM tag will blink.
Press
CFM
to confirm. The secondary LCD will display 600 mV.
Set 600.0 mV on the simulator.
When the reading is stable and close to the selected calibration point,
READY tag will appear and CFM tag will blink.
Press
CFM
to confirm. The secondary LCD will display 1800 mV.
Set 1800.0 mV on the simulator.
When the reading is stable and close to the selected calibration point,
READY tag will appear and CFM tag will blink.
Press
CFM
to confirm. The instrument returns to measurement mode.
Note: If the reading is not close to the selected calibration point,
WRONG tag will blink. Verify calibration condition or contact
your vendor if you can not calibrate.
Press
CAL
in any moment of the calibration process. The instrument
will return to measurement mode. If calibration process is stopped
after 600 mV is confirmed, the 600 mV range is calibrated and
calibration parameters are memorized.