background image

11

Reference
Filling Hole

Reference
Filling Hole

Sensitive

Wire

Reference

Wire

Reference

Junction

Glass

Bulb

Glass
Bulb

Reference
Junction

Reference
Junction

Reference

Junction

Platinum or
Gold tip

Platinum or

Gold tip

Reference
Wire

Reference
Wire

Sensitive
Wire

Glass Body

pH Electrode

Plastic Body

pH Electrode

Plastic Body

ORP Electrode

Glass Body

ORP Electrode

Reference

Wire

ELECTRODE CONDITIONING

AND MAINTENANCE

PREPARATION PROCEDURE

Remove the protective cap.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
removed by "shaking down" the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in 

HI 70300

storage solution for at least one hour.

Note:

 To prevent damage to the electrode, remove the pH

electrode from the solution before turning the meter off.
If the meter is OFF, detach the electrode from the meter
before immersing the electrode in the storage solution.

14

Alkaline Error

TEMPERATURE CORRELATION FOR pH

SENSITIVE GLASS

The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes longer time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 10°C.

Since the resistance of the pH electrode is in the range of 200 Mohm,
the current across the membrane is in the pico Ampere range. Large
currents can disturb the calibration of the electrode for many hours.
For these reasons 

high humidity environments, short circuits and

static discharges

 are detrimental to a stable pH reading.

The pH electrode's life also depends on the temperature. If constantly
used at high temperatures, the electrode life is drastically reduced.

Typical Electrode Life
Ambient Temperature

1- 3 years

90 °C

Less than 4 months

120°C

Less than 1 month

High concentrations of sodium ions interfere with readings in alkaline
solutions; the pH at which the interference starts to be significant
depends upon the composition of the glass. This interference is the
alkaline error and causes the pH to be underestimated. Hanna's
glass formulations have the indicated characteristics.

-20 -20

-10

0 +10+20 +30+40 +50+60+70 +80+90

° C

-10

-10

-10

-10

-10

-10

 Ω

2x10

9

1x10

9

2x10

8

1x10

8

2x10

7

1x10

7

-20   -10      0 

 +

10 

 +

20

+

30

+

40 

+

50 

+

60 

 +

70 

+

80 

 +

90 

   °C

1.0 Mol L

-1

 Na

+

0.1 Mol L

-1

 Na

+

Sodium Ion Correction for the Glass at 20-25°C
Concentration

pH

  Error

13.00
13.50
14.00
12.50
13.00
13.50
14.00

0.10
0.14
0.20
0.10
0.18
0.29
0.40

Содержание HI 22091

Страница 1: ...rintedinROMANIA MAN2209 07 14 Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mail tech h...

Страница 2: ...d by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Technical Service de...

Страница 3: ...correctly All defective items must be returned in the original packing with the supplied accessories 22 n o i t a c i l p p A s e d o r t c e l E e c i u j n o m e L 1 2 B 0 0 1 C F t a e M 2 2 B 1 3...

Страница 4: ...0 2 C F s t c u d o r p y r i a D 6 B 0 0 1 C F B 1 1 9 C F r e t a w y t r i D 7 B 2 1 9 1 I H B 0 1 9 1 I H s n o i s l u m E 8 B 3 5 0 1 I H t n e m n o r i v n E 9 B 0 3 2 1 I H s k s a l F 0 1 B...

Страница 5: ...t 0 g n i s n e d n o c n o n H R 5 9 x a m s n o i s n e m i D m m 9 0 1 x 2 2 2 x 5 3 2 3 4 x 7 8 x 2 9 t h g i e W s b l 9 2 g K 3 1 20 EXTENSION CABLES FOR SCREW TYPE ELECTRODES SCREW TO BNC ADAPT...

Страница 6: ...ottles 50 or 100 mL with the solutions under test on the sample holders Switch the instrument on by pressing the ON OFF switch Press the pH key to display the pH measurement Immerse the electrode tip...

Страница 7: ...ken without temperature compensation Press the mV key to enter the mV mode Allow a few minutes for the reading to stabilize The display will indicate the mV value positive or negative AFTER MEASUREMEN...

Страница 8: ...tion Note If you need to calibrate to NBS standards use pH 6 86 HI 7006 and pH 9 18 HI 7009 instead of pH 7 01 and 10 01 respectively Use a ChecktempC or a glass thermometer as reference Put the beake...

Страница 9: ...temperature see the pH versus temperature chart The pH calibration is now complete 16 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing pretreatment solution 500 mL HI 7092L Oxidizing pretreatment solution...

Страница 10: ...20 mL sachet 25 pcs HI 70007P pH 7 01 buffer solution 20 mL sachet 25 pcs HI 70010P pH 10 01 buffer solution 20 mL sachet 25 pcs HI 7004L pH 4 01 buffer solution 500 mL HI 7006L pH 6 86 buffer solutio...

Страница 11: ...y depends on the temperature The lower the temperature the higher the resistance It takes longer time for the reading to stabilize if the resistance is higher In addition the response time will suffer...

Страница 12: ...sh electrolyte HI 7071 for single junction or HI 7082 for double junction electrodes Allow the electrode to stand upright for 1 hour Follow the Storage Procedure above 13 CLEANING PROCEDURE General So...

Отзывы: