background image

5

GENE
rAL

 D

ESC
ri

PT

iON

 & 

iNTENDED

 U

SE

3.  GENERAL DESCRIPTION & INTENDED USE

HI9810442

 is a professional wireless pH tester, part of Hanna Instruments HALO2 family.

•  The integrated Bluetooth module allows the tester to be connected to a compatible smart device with the Hanna Lab App.
•  The tester has a compact, waterproof casing, and automatic pH calibration at up to three points, or four points when 

used with the Hanna Lab App.

•  Automatically compensated temperature readings are displayed on a large LCD.
•  Accurate and easy to use, the tester is ideal for measurements on flat surfaces, or small volume samples, with the 

specially designed flat sensing tip.

Operating Modes

The 

HI9810442

 can be used as a stand-alone pH tester or connected to the Hanna Lab App.

The Hanna Lab App turns a compatible smart device into a full-featured pH meter. Features include: electrode condition, 

GLP with time-stamp, live-readings, mV resolution, manual temperature compensation, stability criteria, calibration 

reminder, pH (mV) and temperature alarms, tester ID, and data sharing.

Probe Features

The 

HI9810442

 features a non-porous 

glass body

 that is easy to clean and withstands harsh chemicals.

The 

double junction

 design presents a silver-free electrolyte solution interacting with the sample, making the electrode 

less susceptible to clogging and guaranteeing a fast response and stable reading.
Built-in 

temperature sensor

 at the tip of the pH electrode allows for rapid determination of the sample temperature and 

a high-accuracy temperature reading.

4.  FUNCTIONAL DESCRIPTION & LCD DISPLAY

Front & Back view

HI9810442

Leather & Paper

pH T

est

er

LCD

ON/OFF

button

Flat tip

bulb

Maximum 

immersion level

pH Probe

   

SN: XXXXXXXXX
Made in Romania

HI9810442

        

       2

 

1

    

0

4

-

0

1

-

0

A

6

8

7

0

9

2

21  :

CI    

40

B9

A

A 2    :

D I   

C C F   

s n i a t n

Co

R

Setup 

button

Battery cover

LCD display

&

!

#

œ.œ.œ.œ

88.8.8

7

6

5

10

9

8

1
2
3
4

1 Calibration indicator

2 Stability indicator

3 Battery percentage indicator

4 Battery icon

5 Bluetooth icon

6 Temperature compensation

7 Second LCD line

8 Temperature unit

9 First LCD line

10 Measurement unit

Содержание HALO 2 HI9810442

Страница 1: ...INSTRUCTION MANUAL HI9810442 Wireless pH Tester for Leather Paper with built in specialized electrode ...

Страница 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our contact list at www hannainst com Dear Customer...

Страница 3: ...uffer Set 6 pH Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8 10 Care Maintenance 9 Storage 9 11 Warning Error Messages 9 12 Abbreviations 9 13 Accessories 10 Rec...

Страница 4: ...ion pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 02 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers Hanna pH 1 68 4 01 7 01 10 01 or NIST pH 1 68 4 01 6 86 9 18 Temperature compensation Automatic ATC or Manual MTC Electrode Body material Glass Glass Low Temperature LT Junction Ceramic Reference ce...

Страница 5: ... compensation stability criteria calibration reminder pH mV and temperature alarms tester ID and data sharing Probe Features The HI9810442 features a non porous glass body that is easy to clean and withstands harsh chemicals The double junction design presents a silver free electrolyte solution interacting with the sample making the electrode less susceptible to clogging and guaranteeing a fast re...

Страница 6: ...use specified battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup button to enter Setup mode and navigate menu items To exit Setup mode press the Setup button after SEt TIME option is displayed Press the ON OFF button to configure...

Страница 7: ...aring When the tester is in discoverable mode it will appear in the list of Available Devices Withintheapplication tap Connect topairthetesterwiththedevice Allreadingsaretransmitteddirectlytotheapplication Tag a Measurement Once connected to the application the ON OFF button can be pressed to tag the current reading Press the ON OFF button from measurement mode The display shows SEt TAG followed b...

Страница 8: ...de in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE is displayed briefly and the tester will return to measurement mode automatically With 3P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibra...

Страница 9: ... should be kept hydrated When not in use add a few drops of storage solution to the storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water 11 WARNING ERROR MESSAGES WRNG WRNG displayed during calibration Invalid buffer Check the buffer value and use fresh buffer 12 02 23 7 Measured value displayed blink...

Страница 10: ...nvironment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in hazardous environments WARRANTY HI9810442 is warranted for a period of one year against defects in workmanship and materials when used for its intended purpose and maintained according to instructions This warr...

Страница 11: ...12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts de li...

Страница 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810442 Printed in ROMANIA ...

Отзывы: