Refilling the Electrode

•  To remove the electrode sleeve, carefully rotate and slide it off (A). Set it to one side.

Note:

 Handle the probe with care, the pH electrode stem is made of glass.

•  Rinse off any traces of electrolyte gel from sleeve. Soak the electrode tip in 

HI700601

 

General purpose cleaning solution (or 

HI700661

HI700663

HI700664

) for 

20 minutes. Rinse with distilled or deionized water.

•  Refill the reference well with 

HI9071

 Gelled bridge electrolyte (B).

•  Replace the sleeve. Ensure the O-ring is fixed inside the electrode. Rinse off excess gel 

with distilled or deionized water.

•  Shake the probe down gently to eliminate any trapped air bubbles.

•  Soak the probe in 

HI70300

 Electrode storage solution for a minimum of 30 minutes before calibrating.

Storage

To ensure a quick response, the glass tip and junction should be kept hydrated.
When not in use, add a few drops of storage solution to the storage cap. If storage solution is not available, pH 4.01 or 

pH 7.01 buffer can be used.

Note:

 Do not store the electrode in distilled or deionized water.

11.  WARNING & ERROR MESSAGES

----

WRNG

“---- WRNG” displayed during calibration. 

Invalid buffer. 

Check the buffer value and use fresh buffer.

12.02

23.7

Measured value displayed blinking.

Measured value is out of electrode range. 

Clean the electrode to improve condition.

!

8.02

23.7

Battery icon (

!

) displayed blinking.

Battery is below 10 %, replace the battery.

BAtt

DEAD

“BAtt DEAD” and tester turns off.

Depleted battery, replace the battery.

12.  ABBREVIATIONS

ATC 

Automatic Temperature Compensation

GLP 

Good Laboratory Practice

MTC 

Manual Temperature Compensation

NIST  National Institute of Standards and Technology

A

B

Содержание HALO 2 HI9810392

Страница 1: ...INSTRUCTION MANUAL HI9810392 Wireless pH Tester for Chocolate with built in specialized electrode Foodcare ...

Страница 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our contact list at www hannainst com Dear Customer...

Страница 3: ...H Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8 10 Care Maintenance 8 Refilling the Electrode 9 Storage 9 11 Warning Error Messages 9 12 Abbreviations 9 13 Acces...

Страница 4: ...0 to 140 0 F Resolution pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers Hanna pH 1 68 4 01 7 01 10 01 or NIST pH 1 68 4 01 6 86 9 18 Temperature compensation Automatic ATC or Manual MTC Electrode Body material Polyvinylidene Fluoride PVDF Glass Low T...

Страница 5: ...ultraviolet light and fungal growth the PVDF body also has high abrasion resistance and mechanical strength The conical tip allows for easy penetration into solids and semisolids The PVDF outer junction sleeve can be removed and cleaned Once cleaned a small amount of supplied gel electrolyte is added and the junction is refreshed improving the measurement and extending the life of the tester The d...

Страница 6: ...use specified battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup button to enter Setup mode and navigate menu items To exit Setup mode press the Setup button after SEt TIME option is displayed Press the ON OFF button to configure...

Страница 7: ...hen the tester is in discoverable mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measurement Once connected to the application the ON OFF button can be pressed to tag the current reading Press the ON OFF button from measurement mode The display shows SEt TAG fo...

Страница 8: ...de automatically With 3P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 Place the tip of the electrode in the third buffer rinse beaker then in the third calibration buffer Wait until the measurement is stored and the stability indicator disappears...

Страница 9: ... minimum of 30 minutes before calibrating Storage To ensure a quick response the glass tip and junction should be kept hydrated When not in use add a few drops of storage solution to the storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water 11 WARNING ERROR MESSAGES WRNG WRNG displayed during calibrati...

Страница 10: ...using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in hazardous environments WARRANTY HI9810392 is warranted for a period of one year against defects in workmanship and...

Страница 11: ...ovide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interferenc...

Страница 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810392 Printed in ROMANIA ...

Отзывы: