background image

8

CAL

iB

rA

Ti

ON

Stand-Alone Tester (Up to Three-Point Calibration)

1.  Rinse the electrode tip with purified water and blot dry. Then rinse with the buffer being used for calibration.

2.  Press and hold ON/OFF button until “

CAL

 

MODE

” is displayed.

3.  When “

7.01 

USE

” or “

6.86 

USE

” is displayed with “CAL” tag blinking, place the tip of the electrode in the 

correct buffer.

CAL

MODE

   

7.01

USE

4.  When the buffer has been recognized, “

REC

“ is displayed. 

WAIT

” is displayed with the stability indicator (

#

) blinking until the reading is stable.

Wait until the measurement is stored and the stability indicator disappears.

 

7.01

REC

   

#

7..01

WAIT

5.  To save a one-point calibration and return to measurement mode, press the ON/OFF button.

CAL

 

SAVE

” is displayed briefly.

With 2P option selected

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

7.  After the second point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to measurement 

mode automatically.

With 3P option selected

 

6.  Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

7.  Place the tip of the electrode in the third buffer rinse beaker, then in the third calibration buffer. Wait until the 

measurement is stored and the stability indicator disappears. 

8.  After the third point has been stored, “

CAL

 

SAVE

” is displayed briefly and the tester will return to measurement 

mode automatically.

With Hanna Lab App (Up to Four-Point Calibration)

Connect the tester to the Hanna Lab App and follow the calibration procedure.

See the App Help section for the calibration procedure.

Содержание HALO 2 HI9810372

Страница 1: ...INSTRUCTION MANUAL HI9810372 Wireless pH Tester for Skin Scalp with built in specialized electrode ...

Страница 2: ...g a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our contact list at www hannainst com Dear Customer...

Страница 3: ...uffer Set 6 pH Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 Stand Alone Tester Up to Three Point Calibration 8 With Hanna Lab App Up to Four Point Calibration 8 10 Care Maintenance 9 Storage 9 11 Warning Error Messages 9 12 Abbreviations 9 13 Accessories 10 Rec...

Страница 4: ... Temperature 0 0 to 50 0 C 32 0 to 122 0 F Resolution pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers Hanna pH 1 68 4 01 7 01 10 01 or NIST pH 1 68 4 01 6 86 9 18 Temperature compensation Automatic ATC or Manual MTC Electrode Body material Glass Glas...

Страница 5: ...Features The HI9810372 features a non porous glass body that is easy to clean and disinfect The flat glass tip allows for the direct pH measurement of a surface The open junction reference provides for a direct contact with the skin or scalp ensuring a stable measurement The double junction design presents a silver free electrolyte solution interacting with the sample making the electrode less sus...

Страница 6: ...use specified battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Navigation Press the Setup button to enter Setup mode and navigate menu items To exit Setup mode press the Setup button after SEt TIME option is displayed Press the ON OFF button to configure...

Страница 7: ...hen the tester is in discoverable mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measurement Once connected to the application the ON OFF button can be pressed to tag the current reading Press the ON OFF button from measurement mode The display shows SEt TAG fo...

Страница 8: ...de in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE is displayed briefly and the tester will return to measurement mode automatically With 3P option selected 6 Place the tip of the electrode in the second buffer rinse beaker then in the second calibra...

Страница 9: ... should be kept hydrated When not in use add a few drops of storage solution to the storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water 11 WARNING ERROR MESSAGES WRNG WRNG displayed during calibration Invalid buffer Check the buffer value and use fresh buffer 12 02 23 7 Measured value displayed blink...

Страница 10: ...ore using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in hazardous environments WARRANTY HI9810372 is warranted for a period of one year against defects in workmanship...

Страница 11: ...ovide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interferenc...

Страница 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810372 Printed in ROMANIA ...

Отзывы: