Stand-Alone Tester (Up to Three-Point Calibration)
1. Rinse the electrode tip with purified water and blot dry. Then rinse with the buffer being used for calibration.
2. Press and hold ON/OFF button until “
CAL
MODE
” is displayed.
3. When “
7.01
USE
” or “
6.86
USE
” is displayed with “CAL” tag blinking, place the tip of the electrode in the
correct buffer.
CAL
MODE
7.01
USE
4. When the buffer has been recognized, “
REC
“ is displayed.
“
WAIT
” is displayed with the stability indicator (
#
) blinking until the reading is stable. Wait until the measurement is
stored and the stability indicator disappears.
7.01
REC
#
7..01
WAIT
5. To save a one-point calibration and return to measurement mode, press the ON/OFF button.
“
CAL
SAVE
” is displayed briefly.
With 2P option selected
6. Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the
measurement is stored and the stability indicator disappears.
7. After the second point has been stored, “
CAL
SAVE
” is displayed briefly and the tester will return to measurement
mode automatically.
With 3P option selected
6. Place the tip of the electrode in the second buffer rinse beaker, then in the second calibration buffer. Wait until the
measurement is stored and the stability indicator disappears.
7. Place the tip of the electrode in the third buffer rinse beaker, then in the third calibration buffer. Wait until the
measurement is stored and the stability indicator disappears.
8. After the third point has been stored, “
CAL
SAVE
” is displayed briefly and the tester will return to measurement
mode automatically.
With Hanna Lab App (Up to Four-Point Calibration)
Connect the tester to the Hanna Lab App and follow the calibration procedure.
See the App Help section for the calibration procedure.
10. SUSHI RICE MEASUREMENT
The application specific electrode design allows for
direct measurement of sushi rice
and
measurement of rice slurries
.
Preparation
• Calibrate the tester prior to measurement. Refer to calibration section for details.
• Loosen the threaded top collar counterclockwise before removing the probe from storage cap.
After measurement, place the probe tip into the cap and gently tighten the threaded collar.
• Do not be alarmed if salt deposits or storage solution are present. Rinse the electrode with purified water.
• If the glass and / or junction are dry, soak the electrode in Storage solution (or pH 4.01 or pH 7.01 buffer solution)
for a minimum of 30 minutes.
• After conditioning, rinse with water and blot dry.