Hanna Instruments Foodcare HI981034 Скачать руководство пользователя страница 2

If pH 4.01 buffer 
solution is used 
as the first point

 

the value of the 
buffer is recognized 
and displayed 
with the blinking 
stability indicator.

If pH 7.01 buffer solution is used as 
the first point

 the buffer is recognized 

with the blinking stability indicator.
When the reading is stable, the 
stability indicator will disappear 
and pH 7.01 will be calibrated. 

If pH 7.01 is the only calibration point,

 

finish one-point procedure at right. 

If using pH 4.01 as a second point,

 

continue two-point procedure at right.

For one-point calibration using pH 4.01 buffer go to procedure B 

For one or two-point calibration using pH 7.01 buffer go to procedure A

A

B

When the 
reading is stable, 
the stability 
indicator will 
disappear.“Sto” 
will be displayed 
when the 
calibration 
is saved.

Meter will 
exit to 
measurement 
mode and the 
calibration 
tag will be 
displayed.

Meter will exit 
to measurement 
mode and the 
calibration 
tag will be 
displayed.

Tw

o-

Po

in

t

Use pH 4.01 to 
perform a two-
point calibration. 
The value is 
automatically 
recognized and 
displayed with the 
blinking stability 
indicator.

When the reading 
is stable, the 
stability indicator 
will disappear. 
“Sto” will be 
displayed when 
the calibration 
is saved.

Meter will exit 
to measurement 
mode and the 
calibration 
tags will be 
displayed

.

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using this instrument. 
For technical support, contact your local Hanna Instruments Office or email us 
at [email protected].
To find your local Hanna Instruments Office or for additional information on 
Hanna Instruments products, visit www.hannainst.com

Meter Calibration

From measurement 
mode, press and 
hold the ON/OFF 
button until “CAL” 
is displayed.

When “7.01” blinks 
on the display, place 
the tip of the probe 
into a pH 7.01 or 
4.01 buffer solution.

On

e-

Po

in

t

One-Point Calibration with pH 4.01

B

A

One or Two-Point Calibration with pH 7.01

pH 4.01 will then 
blink on the display. 
Ignore it and press 
ON/OFF button.

Operation

Press the ON/OFF button to turn the meter on. All tags 
will be displayed.

The meter will go into measurement mode: current 
reading and calibrated buffers are displayed.

Preliminary Examination

Remove the meter from the packing material and examine it carefully to make sure that 
no damage has occurred during shipment. If noticeable damage is evident, contact your 
local Hanna Instruments Office.
Each meter is supplied with:

• pH 4.01 buffer solution liquid sachet (2 pcs.)
• pH 7.01 buffer solution liquid sachet (2 pcs.)
• Cleaning solution for milk deposits ( 2 pcs.)
• Electrode storage solution, 13 mL dropper
• Instruction manual
• Quality certificate

Note: Save all packing material until you are sure that the instrument functions 

correctly. All defective items must be returned in the original packaging with the 

supplied accessories.

Meter Overview

Preparation:

The pH electrode is shipped with a protective cap containing storage solution. Before 
using the meter, 

remove the protective cap

 and condition the electrode by soaking the 

tip (bottom 4 cm (1.5”)) in pH 7.01 buffer solution for several minutes. Then follow the 
calibration procedure.

•  Do not be alarmed if white crystals appear around the cap. This is normal with pH 

electrodes and they dissolve when rinsed with water.

•  Turn the meter on by pressing ON/OFF button.
•  Remove the protective cap and immerse the tip of the electrode in the sample 

to be tested.

NEVER IMMERSE THE ELECTRODE OVER THE MAXIMUM IMMERSION LEVEL.

•  Stir gently and wait for a stable reading.

 

•  For best results, recalibrate periodically.
•  After use, rinse the electrode with water and store it with a few drops of storage 

solution in the protective cap.

•  Reattach the protective cap after each use.

DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.

Intended use

The measurement of pH in milk is important in testing for impurities, spoilage, and signs 
of mastitis infection. While there are a number of factors that affect the composition of 
milk, pH measurements can help producers understand what might be causing certain 
compositional changes. pH measurement is commonly performed at various points in a 
milk processing plant.
Milk with pH values higher than pH 6.7 potentially indicate that the milk may have come 
from cows with a mastitis infection. Mastitis is an ever-present challenge with dairy milking 
cows. When infected, the cow's immune system releases histamine and other compounds 
in response to the infection. There is a resulting increase in permeability of endothelial and 
epithelial cell layers, allowing blood components to pass through a paracellular pathway. 
Sinceblood plasma is slightly alkaline, the resulting pH of milk will be higher than normal. 
Typically milk producers can perform a somatic cell count to detect a mastitis infection, but a 
pH measurement offers a quick way to screen for infection.
Understanding the pH of raw milk can also help producers optimize their processing 
techniques. For example, in operations that use Ultra High Temperature (UHT) processing, 
even small variations from pH 6.7 can affect the time required for pasteurization and the 
stability of the milk after treatment.

Probe Features

Low Temperature Glass Tip and Open Junction.

The glass tip uses a special low temperature (LT) glass 

formulation. This is beneficial since many food products are at 

low temperatue. The open junction design consist of a solid gel 

interface between the sample and the internal reference wire. 

This makes it impermeable to clogging after measurements. 

Conical Glass Tip

The conical shaped tip design allows for penetration into solids, semi-solids, and emulsions 

for the direct measurement of pH in samples such as milk and yogurt.

Liquid Crystal 
Display

ON/OFF 
Button

Maximum 
Immersion 
Level

pH Probe

Calibration 

tags

Battery 

indicator

Unstable 

tag

Measurement 

unit

Digits

Recommended 
Immersion 
Level

Range

0.0 to 14.0 pH

Resolution

0.1 pH

Accuracy 

±0.2 pH @25 °C/77 °F

Calibration

Automatic, one or two-point

Electrode

Built-in probe for specific application

Battery Type

CR2032 Li-ion

Battery Life

Approximately 1000 hours of continuous use

Auto-off

8 minutes, 60 minutes or can be disabled

Environment

0 to 50 °C (32 to 122 °F); RH 95% max

Dimensions

51 x 159 x 21 mm (2 x 6.3 x 0.9“) 

Weight

50 g (1.8 oz.)

Specifications

"Sto" will be 
displayed when 
the calibration 
is saved.

Отзывы: