Setup/Installation
14
In Setup mode, choose log parameter, press
MODIFY
then use the
u
arrow to
select between Interval, Manual, or Stability. When Interval is displayed, use the
p
and
q
arrows to select the setting for the timed interval. When Stability is
displayed, use the
p
and
q
to select the measurement stability setting.
A complete set of GLP information including date, time, range selection,
temperature reading, calibration information and probe serial number is stored
with each log made.
Interval Logging
Select Interval and sampling period in the SETUP menu (Not available in Basic
mode). To start Interval logging, press the
LOG
key while the instrument is in
measurement mode.
A “PLEASE WAIT” message will be displayed followed
by the number of free spaces. During active interval
logging, lot information is displayed on the third
LCD line. The line indicates in which lot the data will
be placed and keeps track of the number of logged
records taken. The “LOG“ tag is continuously on during
active logging.
Pressing
u
during an interval log will display the
number of logs available.
Pressing the
LOG
key again will stop the Interval
logging session. The “LOG STOPPED“ message will be
displayed for a few seconds.
If a sensor failure occurs during interval logging, the
message “OUT OF SPEC.” will alternate with logging
information.
Manual Logging
Select Manual in the SETUP menu. To initiate a
Manual log, press the
LOG
key while the instrument
is in measurement. The “PLEASE WAIT” screen will
be displayed briefl y followed by a screen indicating
the measurement has been saved and then a screen
indicating the log record number.
The “LOG” tag will be displayed on all 3 screens.
“PLEASE WAIT”
“SAVED” with the log record number
“FREE” with the number of free spaces available
Logging
Function