Hanna Instruments edge DO Скачать руководство пользователя страница 28

Maintenance

28

The DO probe body is made of PEI.
Use the protective cap when the probe is not in use. To replace the membrane or 
refi ll with electrolyte, proceed as follows:

For  a  new  probe,  remove  the  protective  shipping  tube  by  gently 
twisting and pulling it off  the body of the probe (see fi g. 1).

If the membrane was previously installed, unscrew the membrane 
cap by turning it clockwise (see fi g. 2).

The new membrane cap should be rinsed with electrolyte solution. 
Refi ll with clean electrolyte solution.

Gently  tap  the  sides  of  the  membrane  cap  to  ensure  that  no  air 
bubbles  remain  trapped.  Do  not  tap  the  bottom  directly  with  your 
fi nger, as this will damage the membrane.

Make sure that the rubber O-ring sits properly inside the membrane 
cap. With the sensor facing down, slowly screw the membrane cap 
counterclockwise. Some electrolyte will overfl ow.

The  platinum  cathode  (DO  Probe  Diagram,  page  21)  should  always  be  bright 
and  untarnished.  If  it  is  tarnished  or  stained,  the  cathode  should  be  cleaned. 
You  can  use  a  clean  lint-free  cardboard  or  cloth.  Rub  the  cathode  very  gently 
side  to  side  4-5  times.  This  will  be  enough  to  polish  and  remove  any  stains 
without damaging the platinum tip. Afterwards, rinse the probe with deionized 
or distilled water and install a new membrane cap using fresh electrolyte and 
follow DO Probe Preparation page 28.

Important

In  order  to  have  accurate  and  stable  measurements,  it  is  important  that  the 
membrane  surface  is  in  perfect  condition.  This  semipermeable  membrane 
isolates  the  sensor  elements  from  the  environment  but  allows  oxygen  to 
enter. If any dirt is observed on the membrane, rinse carefully with distilled or 
deionized water. If imperfections still exist, or any damage is evident (such as 
wrinkles or tears-holes), the membrane should be replaced. Make sure that the 
O-Ring sits properly in the membrane cap.

Dissolved 

Oxygen 

Probe 

Maintenance

Maintenance

Содержание edge DO

Страница 1: ...DO...

Страница 2: ...of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Included 3 Safety Measures 3 Description 4 Diagram 4 8 Product Diagram 4 5 Probe Diagram 6 Keypad Function 7 Guide to Ind...

Страница 3: ...rings for membrane caps Before using this product make sure that it is entirely suitable for your specific application and for the environment in which it is used Operation of this instrument may caus...

Страница 4: ...libration measurement data logging and transfer of data to a USB thumb drive or computer Every feature and measurement detail is designed to give you an edge in measurement technology edgeDO is versat...

Страница 5: ...2 Capacitive Touch Keypad 3 3 mm jack input for edgeDO digital probes 4 Top mounted ON OFF button 5 Micro USB device connection for power or PC interface 6 Standard USB host connection for data transf...

Страница 6: ...sensor recognition Store calibration specific data from the last calibration Are built with materials suitable for use in chemical analysis Have integrated temperature measurement Incorporate a 3 mm...

Страница 7: ...TUP CLR Used to enter exit SETUP mode During calibration used to clear previous calibration data In log RCL used to clear log records 5 Used to scroll through SETUP menu Used to change selection when...

Страница 8: ...nt 10 Temperature units 11 Temp Compensation status 12 Measurement line 13 Measurement units 14 Stability Indicator The third line of the LCD 9 is a dedicated message line During measurement the user...

Страница 9: ...d for the measurement you will be making 6 Calibrate the sensor probe You are now ready for measurements Bench Cradle Setup Insert electrode holder arm into the post on the pivoting base Connect the p...

Страница 10: ...t the top Plug the 5 VDC adapter into the power supply socket or by connecting directly to a PC Note edgeDO is supplied with a rechargeable battery inside which provides about 8 hours of continuous us...

Страница 11: ...ata will be exported to the PC LOG ON EDGE or EXPORT TO PC LOG ON EDGE Log Select log type to be used from 3 types of logging Manual log on demand Manual log on stability 3 types of stability criteria...

Страница 12: ...lash Format Only seen when log errors are present Permits formatting the flash drive On or Off OFF Message Transition User may choose how messages are displayed on third LCD line of display Word scrol...

Страница 13: ...urement type New records made on different days are stored in the same Manual lot Manual Stability log on demand A log on demand that is made each time LOG key is pressed and the stability criteria is...

Страница 14: ...top the Interval logging session The LOG STOPPED message will be displayed for a few seconds If a sensor failure occurs during interval logging the message OUT OF SPEC will alternate with logging info...

Страница 15: ...pressing LOG again When the stability selected criteria has been met a SAVED message will be displayed followed by a screen indicating how much log space is available The LOG tag will be displayed on...

Страница 16: ...te Press CLR The instrument will display CLEAR MANUAL for Manual Records CLEAR STAB for Stability Records For Interval lots the message CLEAR followed by the selected lot will be displayed with CFM ta...

Страница 17: ...were deleted The total number of lots that can be saved is 100 If some are deleted for example 1 50 fifty additional logs may be stored These will be numbered 101 150 The lots are allocated successiv...

Страница 18: ...followed by EXPORT Press CFM to export the selected record or lot If CFM is not pressed in 10 seconds the USB host will become inactive The meter will display the percentage of export The export perc...

Страница 19: ...ote Western Europe ISO 88859 1 character set and English language are suggested settings Other files may be visible depending upon computer settings All files stored will appear in this folder Adjust...

Страница 20: ...the cardboard shipping tube used to protect the probe during shipping Save the tube should the probe be stored dry again 2 Open membrane package and remove one O ring and one membrane cap 3 Rinse the...

Страница 21: ...reaches an equilibrium with the surrounding solution Whenever measurements are taken with a non polarized probe the measurement will be drifty and inaccurate The measurement will jump when the probe...

Страница 22: ...the most accurate calibration and concentration measurement 500 400 300 200 100 0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400 2500 2...

Страница 23: ...18 less compared to fresh water at 25 C By entering the approximate salinity value the calibration and subsequent concentration measurement will be compensated to display the correct oxygen concentrat...

Страница 24: ...leave calibration after the first point The instrument will display SAVING message and it will return to measurement mode memorizing the slope calibration data Zero Calibration PressCALorcontinuewith...

Страница 25: ...standard Altitude and Salinity factors time and date of the last calibration the expired calibration information and the probe serial number This information is also included with all logged data Last...

Страница 26: ...2 DAYS AGO The probe serial number together with the current reading Make sure that the probe is polarized calibrated and the protective cap has been removed Rinse probe Submerse the probe in the samp...

Страница 27: ...e between the temperature at which the probe was stored and the temperature of the sample the longer the time will be If the DO temperature exceeds the limits of the probe the message PROBE OUT OF SPE...

Страница 28: ...ow The platinum cathode DO Probe Diagram page 21 should always be bright and untarnished If it is tarnished or stained the cathode should be cleaned You can use a clean lint free cardboard or cloth Ru...

Страница 29: ...t and clean if necessary Install new cap fresh electrolyte with no bubbles permit longer polarization Stir or increase flow rate The meter does not measure temperature is displayed on second LCD line...

Страница 30: ...g L 1 g L Temperature Compensation 0 0 to 50 0 C 32 0 to 122 F DO Probe HI 764080 Log Feature Up to 1000 records organized in Manual Log on Demand Max 200 logs Manual Log on Stability Max 200 logs Int...

Страница 31: ...30 mL HI 764080 Spare DO probe HI 764080A P 5 spare membranes Other Accessories HI 75110 220U Voltage adapter from 115 VAC to 5 VDC USA plug HI 75110 220E Voltage adapter from 230 VAC to 5 VDC Europe...

Страница 32: ...l Hanna Office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges i...

Отзывы: