background image

WARRANTY

These meters are guaranteed for one year against de-
fects in workmanship and materials when used for its
intended purpose and maintained according to instructions.
Probes are guaranteed for a period of six months. This
warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of
prescribed maintenance are not covered. If service is re-
quired, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number,
date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will
be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Returned
Goods Authorization Number from the Customer Service
department and then send it with shipment costs prepaid.
When shipping any instrument, make sure it is properly
packaged for complete protection.

ISTCKTDIP R3      05/05

BATTERY REPLACEMENT

When 

Checktemp

®

-Dip

 cannot be switched on or

the display fades, remove the two screws, remove
the battery cover and change the 1.5V AAA bat-
tery, while paying attention to its polarity.
Replace the cover and tighten the two screws.
Battery should only be replaced in a safe area
and using an alkaline 1.5V AAA type battery.

OPERATION

• Turn the 

Checktemp

®

-Dip

 on and dip the probe in

the sample to be measured.

• Do NOT wrench the probe. If the cable is dam-

aged, the meter may have to be replaced.

• Wait a few seconds for the display to stabilize.
• After use, always switch the meter off
• Variations in readings or a fading display can

indicate a low battery level (see “Battery Re-
placement” section for details).

Checktemp

®

-Dip

  should not be used for continu-

ous measurement at high temperatures. In fact,
even though 

Checktemp

®

-Dip

  can reach 100ºC

(212°F), prolonged measurements at high tem-
perature will reduce the sensor life dramatically.
The ideal operating temperature is in the -20 to
50ºC (-4 to 120ºF) range.

CALIBRATION CHECK SWITCH

Move the Cal-check

®

 switch on the back of the

meter to the “Test” position.
The display should show 0.0°C (±0.5°C) or 32ºF
(±0.9°F).
Contact your dealer or the nearest Hanna Service
Center for recalibration, if necessary.
Move the Cal-check switch back to the “Run” po-
sition, and the display will return to normal mea-
surement mode.

Recommendations for Users

Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residential
areas could cause interference to radio and TV equipment. Any variation
introduced by the user to the supplied equipment may degrade the
instrument's EMC performance. To avoid electrical shocks, do not use this
instrument when voltage at the measurement surface exceeds 24 Vac or
60 Vdc. To avoid damage or burns, do not perform any measurement in
microwave ovens.