background image

Battery

ON/OFF button

Liquid Crystal Display

Maximum Immersion Level

Thank you for choosing a Hanna Instruments product. Please read 
this instruction manual carefully before using the instrument. For 
additional information, visit www.hannainst.com or e-mail us at 
[email protected].

Specifi cations

Range

0.00 to 14.00 pH

Resolution

0.01 pH

Accuracy

±0.1 pH

Calibration

Automatic, 2 points

Electrode

HI 1271

 (included)

Environment

0 to 50 ºC (32 to 122 ºF); RH 95% max

Battery Type

1 x CR2032 Li ion

Battery Life

approx. 1000 hours of continuous use

Auto Off 

OFF, 8 or 60 minutes

Dimensions

50 x 174 x 21 mm (2 x 6.8 x 0.9“)

Weight

50 g (1.8 oz.)

Thank You!

Operation

Replace the battery when the battery symbol

 

on the LCD blinks or 

the meter cannot be switched on. When the battery is low “

Erb

“ 

message appears and the meter turns off  automatically.
Remove the battery cover on the back of the meter. Replace the 
battery with a new CR2032, while paying attention to its polarity.

Note:

  Batteries should only be replaced in a safe area using the 

battery type specifi ed in this instruction manual. Old batteries 
should be disposed in accordance with local regulations.

Battery Replacement

• For best results, calibrate frequently. 
• From measurement mode, press and hold 

ON/OFF

 button until 

CAL

” message appears on display.

• Release the button and dip the tip of the electrode (bottom 

4 cm/1½”) in 7.01, 4.01 or 10.01 pH buff er solution at room 
temperature.

• On the display 7.01 pH will appear blinking.
• When the buff er is recognized the buff er value will appear and 

“ “ will blink.

• The calibration is automatically performed.
• If the fi rst confi rmed buff er is 4.01 or 10.01 pH, the calibration 

is done “

Sto

“ will appear and the checker returns to normal 

measurement mode.

• If the fi rst confi rmed buff er is 7.01 pH, the checker asks for second 

calibration point.

• On the display 4.01 pH will appear blinking.
• By pressing 

ON/OFF

 button, the calibration is done only in 7.01 pH 

buff er.  “

Sto

“ will appear and the checker returns to normal 

measurement mode.

• For second calibration point, rinse the electrode with water and 

dip it in pH 4.01 (or pH 10.01) buff er solution.

• Once the calibration has been automatically performed “

Sto

“ will 

appear and the checker will return to normal measurement mode 
with the calibrated buff er tags ON.

ALWAYS CALIBRATE USING FRESH BUFFER SOLUTIONS.

CLEAR CALIBRATION

• After entering in calibration mode, press and hold 

ON/OFF

 button 

until “

Clr

” message appears on display.

• The checker will return to measurement mode with default 

calibration.

Notes:
• If “

Err

“ message appears, check if the meter is in the correct buff er 

or clean the electrode using 

HI 7061M

 cleaning solution.

• Press and hold the 

ON/OFF

 button until “

d60

” or “

d08

” or “

d--

” 

appears. Press the 

ON/OFF

 button to cycle through the following 

Auto Shut Off  options: 8 minutes (“

d08

”, default value), 60 minutes 

(“

d60

”), OFF (“

d--

”). Press and hold the 

ON/OFF

 button to save 

and continue to normal operation.

Meter Calibration

Checker Plus® is guaranteed for a period of one year after date of purchase 
against defects in workmanship and materials when used for their intended 
purpose and maintained according to instructions. This warranty is limited to 
repair or replacement free of charge. Damage due to accidents, misuse, tampering 
or lack of prescribed maintenance is not covered. If service is required, contact 
your local Hanna Offi

  ce. If under warranty, report the model number, date of 

purchase, serial number and the nature of the problem. If the repair is not 
covered by the warranty, you will be notifi ed of the charges incurred. If the 
instrument is to be returned to Hanna Instruments, fi rst obtain a Returned 
Goods Authorization number from the Technical Service department and 
then send it with shipping costs prepaid. When shipping any instrument, 
make sure it is properly packed for complete protection.

Before using Hanna Products, make sure that they are entirely suitable for 
your specifi c application and for the environment in which they are used. 
Operation of these instruments may cause unacceptable interferences to 
other electronic equipment, thus requiring the operator to take all necessary 
steps to correct such interferences. Any variation introduced by the user to 
the supplied equipment may degrade the instrument’s EMC performance. To 
avoid damages or burns, do not put the instrument in microwave oven. For 
yours and the instrument safety do not use or store the instrument in 
hazardous environments.

All rights are reserved. Reproduction in whole or in part is prohibited without 
the written consent of the copyright owner, Hanna Instruments Inc., 
Woonsocket, Rhode Island, 02895, USA.

IST98100

03/15

Warranty

Recommendations for Users

 

The pH electrode is shipped dry. Before using the Checker Plus

®

remove the protective cap and condition the electrode by soaking 
the tip (bottom 4 cm/1½”) in pH 7.01 buff er solution for a couple of 
hours. Then follow the calibration procedure. 
• Do not be alarmed if white crystals appear around the cap. This 

is normal with pH electrodes and they dissolve when rinsed with 
water.

• Connect the electrode to the meter.
• Turn the Checker Plus

®

 on by pressing 

ON/OFF

 button.

• Remove the protective cap and immerse the tip of the electrode 

(bottom 4cm/1½”) in the sample to be tested.

• Stir gently and wait for stable reading.
NEVER IMMERSE THE ELECTRODE OVER THE MAXIMUM IMMERSION 
LEVEL. THE CONNECTOR MUST BE ALWAYS CLEAN AND DRY.
• After use, rinse the electrode with water and store it with a few 

drops of 

HI 70300

 storage solution in the protective cap.

DO NOT USE DISTILLED OR DEIONIZED 
WATER FOR STORAGE PURPOSES.
•  Always replace the protective cap after 

use.

4 cm
1.5”

Recommended
Immersion Level

pH Probe