Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows
“Add”
,
“C1”
with “
Press
” blinking,
the meter is ready.
•
Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
•
Press the button. When the display shows
“
Add
”, “
C.2
” with “
Press
” blinking the meter
is zeroed.
•
Remove the cuvette, from the meter and
unscrew the cap. Add the content of one packet
of
HI 716-0
reagent. Replace the cap and
shake gently for 20 seconds to dissolve most
of the reagent. Place the cuvette back into
the meter.
•
Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 2 minutes and
30 seconds and press the button.
•
The instrument directly displays the
concentration of bromine in ppm. The meter
automatically turns off after 10 minutes.
10 mL
REAGENT SETS
HI 716-25
Reagents for 25 Bromine tests
OTHER ACCESSORIES
HI 716-11
Bromine Certified Standard Kit
HI 731318
Cloth for wiping cuvettes (4 pcs.)
HI 731321
Glass cuvettes (4 pcs.)
HI 731225
Cuvette black cap for checker HC (4 pcs.)
HI 731353
Cuvette seal cap for checker HC (4 pcs.)
HI 740028
1.5V AAA batteries (4 pcs.)
HI 93703-50
Cuvettes cleaning solution (230 mL)
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments,
thus requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
w w w . h a n n a i n s t . c o m
Hanna Instruments
reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
IST716 03/13
Measurement procedure:
Accessories:
Tips for an accurate measurement
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “
bAd
” then ”
bAt
” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the
negative end first.
• Insert the battery cover and replace the screw with a screwdriver.
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside,
and completely free of fingerprints, oil and dirt. Wipe it thoroughly with
HI 731318
or a lint-free cloth prior to insertion.
• Shaking the cuvette can generate bubbles, causing lower readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
• Do not let the reacted sample stand for too long after reagent is added, as accuracy
will be affected.
• After the reading it is important to immediately discard the sample, otherwise the
glass might become permanently stained.