Hanna Instruments Ammonia Medium Range Скачать руководство пользователя страница 2

10 mL

• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it must be dry

outside, and completely free of fingerprints, oil and dirt. Wipe it thoroughly with

HI 731318

 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher readings. To obtain

accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.

• Do not let the reacted sample stand for too long after reagent is added, as

accuracy will be affected.

• After the reading it is important to immediately discard the sample, otherwise

the glass might become permanently stained.

REAGENT SETS
HI 715-25

Reagent set for 25 Ammonia Medium Range tests

OTHER ACCESSORIES
HI 715-11

Ammonia MR Certified Standard Kit

HI 731318

Cloth for wiping cuvettes (4 pcs.)

HI 731321

Glass cuvettes (4 pcs.)

HI 731225

Cuvette black cap for checker HC (4 pcs.)

HI 731353

Cuvette seal cap for checker HC (4 pcs.)

HI 740028

1.5V AAA batteries (4 pcs.)

HI 93703-50

Cuvette cleaning solution (230 mL)

Recommendations for Users

Before using this product, make sure that it is entirely suitable for your specific application and for the
environment in which it is used.
Operation of this instrument may cause unacceptable interferences to other electronic equipment, thus
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damage or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.

For additional information, contact your

dealer or the nearest

Hanna Customer Service Center.

To find a Hanna Office in your area,

visit our web site:

w w w . h a n n a i n s t . c o m

Hanna Instruments 

reserves the right to modify the design, construction or

appearance of its products without advance notice.

Battery management

To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead
the instrument will display “

bAd

” then ”

bAt

” for 1 second and then turns off.

To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.

• Remove the battery from its location and replace it with a new one, inserting the

negative end first.

• Insert the battery cover and replace the screw with a screwdriver.

IST715   03/13

Measurement procedure:

Accessories:

Tips for an accurate measurement

Turn the meter on by pressing the button.
All segments will be displayed. When the
display shows “

Add

”, “

C.1

” with “

Press

blinking, the meter is ready.

Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.

Press the button. When the display shows

Add

”, “

C.2

” with “

Press

” blinking the

meter is zeroed.

Remove the cuvette from the meter and
unscrew the cap. Add 4 drops of 

HI 715A-0

reagent A. Replace the cap and swirl the
solution.

Unscrew the cap and add 4 drops of 

HI 715B-0

reagent B. Replace the cap and swirl the
solution. Place the cuvette back into the
meter.

Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 3 minutes and
30 seconds and press the button.

The instrument displays the results in mg/L
(ppm) of ammonia nitrogen (NH

3

-N). To

convert the reading to ppm of ammonia
(NH

3

), multiply the reading by factor of

1.214. The meter automatically turns off
after 10 minutes.

[email protected]

www.ponpe.com

Отзывы: