13

• Place the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press 

ZERO/CFM 

and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• After a few seconds, the display will show

“-0.0-”

. The meter is now zeroed and

ready for measurement. Remove the cuvette.

• Add one packet of the 

HI 93728-0

reagent.

• Replace the cap and immediately shake

vigorously for exactly 10 seconds by
moving the cuvette up and down. Continue
to mix by inverting the cuvette gently
and slowly for 50 seconds, while taking
care not to induce air bubbles. A deposit
could remain but it dose not affect the
measurement. Time and way of shaking
could sensitively affect the measurement.

• Replace the cuvette into the holder and

ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold 

READ/ /TIMER

 for three

seconds. The display will show the
countdown prior to measurement. An
audible “beep” indicates the end of the
countdown period.

Содержание 96728

Страница 1: ...1 Instruction Manual HI 96728 Nitrate ISM...

Страница 2: ...RY MANAGEMENT 21 BATTERY REPLACEMENT 22 ACCESSORIES 22 WARRANTY 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use...

Страница 3: ...ing shipment please notify your Dealer Each HI 96728 Ion Selective Meter is supplied complete with Two Sample Cuvettes and Caps 9V Battery Instrument quality certificate Instruction Manual Note Save a...

Страница 4: ...m dust or dirt by a transparent cup This makes the instruments fulfill field applications Display messages aid the user in routine operation The meter has an auto shut off feature that will turn off t...

Страница 5: ...se in calibration mode with last reading reminder Dimensions 192 x 104 x 69 mm 7 6 x 4 1 x 2 7 Weight 360 g 12 7 oz REQUIRED REAGENTS Code Description Quantity test HI 93728 0 Powder Reagent 1 packet...

Страница 6: ...ion coefficient at wavelength c molar concentration of the substance d optical path through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the ot...

Страница 7: ...coming value into the desired measuring unit and to display it on the LCD The measurement process is carried out in two phases first the meter is zeroed and then the actual measurement is performed Th...

Страница 8: ...de press to edit the date and time OPERATING MODES Measurement mode default operation mode enables both validation and measurement Calibration mode may be entered by keeping CAL CHECK pressed for thre...

Страница 9: ...ht to perform a measurement Please check the preparation of the zero cuvette No Light The instrument cannot adjust the light level Please check that the sample does not contain any debris Inverted cuv...

Страница 10: ...igh The standard reading is higher than expected Cap error Appears when external light enters in the analysis cell Assure that the cuvette cap is present Cooling lamp The instrument waits for the lamp...

Страница 11: ...ASUREMENT It is important that the sample does not contain any debris This would corrupt the reading Each time the cuvette is used the cap must be tightened to the same degree Whenever the cuvette is...

Страница 12: ...trument by removing the dust protection sleeve from the instrument cuvette holder Place the battery in the instrument as described in the BATTERY REPLACEMENT chapter Place the instrument on a flat tab...

Страница 13: ...p and immediately shake vigorously for exactly 10 seconds by moving the cuvette up and down Continue to mix by inverting the cuvette gently and slowly for 50 seconds while taking care not to induce ai...

Страница 14: ...erference may be caused by Ammonia and amines as urea and primary aliphatic amines Chloride above 100 mg L negative interference Chlorine above 2 mg L positive interference Copper it must be absent Ir...

Страница 15: ...eroed first Place the CAL CHECKTM Standard HI 96728 11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press ZERO CFM and the lamp cuvette...

Страница 16: ...ent the display will show the validation standard value The reading should be within specifications as reported in the CAL CHECKTM Standard Certificate If the value is found out of the specifications...

Страница 17: ...isplays dashes the meter is ready Press and hold CAL CHECK for three seconds to enter calibration mode The display will show CAL during calibration procedure The blinking ZERO asks for instrument zero...

Страница 18: ...display depending on the measurement phase After measurement the instrument will show for three seconds the CAL CHECKTM Standard value Note If the display shows STD HIGH the standard value was too hi...

Страница 19: ...ed the month number is automatically increased When the correct month has been set press ZERO CFM or READ TIMER to confirm Now the display will show the day blinking Press GLP to edit the desired day...

Страница 20: ...ppear on the main display and the instrument returns to measurement mode after three seconds It is possible to delete the calibration and restore factory calibration Press GLP to enter GLP mode In GLP...

Страница 21: ...easurement was displayed before auto shut off the value is displayed when the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 7...

Страница 22: ...it clockwise to close To replace the instrument s battery follow the steps ACCESORIES REAGENT SET HI93728 01 Reagentsfor100tests HI93728 03 Reagentsfor300tests OTHER ACCESORIES HI96728 11 CAL CHECKTM...

Страница 23: ...tained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If serv...

Страница 24: ...24 MAN96728 01 11 PrintedinEUROPE ROMANIA Local Sales and Customer Service Office...

Отзывы: