background image

Before Operation

After Operation

Check the rotor is firmly tightened onto

the motor shaft.

Check the Rotor Locking

After relocating the device or loading the rotor, 

balance them by using a leveler. If necessary, level 

the device by adjusting the‘Leveling foot’on the 

bottom of the device.

Balancing the Centrifuge and Rotor

Cautions For Use

After cleaning, lubricate

bucket grooves, pivots of

swing-bucket rotors and

threads of the angle rotors

using the lubricant.

Wipe up the moisture in the

chamber with a dry cloth after

use. Keep the lid open when not 

in use.

Clean the chamber and 

rotor by wiping with a cloth 

soaked in mild detergent and

dry completely.

Care of Swing-Out Rotor

Keep the Chamber Dry

Cleaning

Check the rotor, buckets and tubes are

free of scratch and crack before operation.

Check the Accessories for Abnormalities

Load tubes symmetrically into the

rotor with identical amount.

Load Tubes Symmetrically

A-15-24    
A-50c-10    
A-50-10  

S-mw-2
S-15-6 
S-50-4
S-100-4

24 x 15 mL / Max. 5k
10 x 50 mL Conical / Max. 5k
10 x 50 mL / Max. 4k

2 x 4 MTP / Max. 4k
6 x 2 x 15 mL / Max. 4.4k
4 x 50 mL / Max. 5k
4 x 100 mL / Max. 4.5k

Rotor

Electrical Safety

Switch off the device after

use. Unplug the power plug

before cleaning or unused for 

a long period of time.

Check the wall outlet

is earth grounded,

surge-protected andcan

supply enough power.

Switch Off the Device

Grounded Outlet

Check the Line Voltage

Maintenance

If the centrifuge lid cannot be opened, insert

the provided T-wrench into the emergency

release hole on the left side of the device and

turn it clockwise. (Before activating

emergency open, switch off the device and

wait for the rotor to stop.)

Emergency Lid Open

Swing-out Rotors

Fixed-angle Rotors

Connect the device to voltage

sources which correspond to

the electrical requirements on

the label attached to the device.

900VA 
220V  50/60Hz  
(110V Optional)

Keep a safety clearance of

30 cm around the centrifuge

during operation.

Empty the Condensate

Keep a Safety Clearance

Empty the condensate

regularly by removing the 

drain cap on the left side of

the device

30 cm

30 cm

30 cm

HI-QG-T05R(E)(Rev.1)

Quick Guide

T05R

Cat.  No. :  T05R

Order & Inquiry l +82-2-3472-0727  l [email protected] l www.ihanil.com

Service  l +82-2-3452-8966 l [email protected]