Before Operation
After Operation
Check the rotor is firmly tightened onto
the motor shaft.
Check the Rotor Locking
After relocating the device or loading the rotor,
balance them by using a leveler. If necessary, level
the device by adjusting the‘Leveling foot’on the
bottom of the device.
Balancing the Centrifuge and Rotor
Cautions For Use
After cleaning, lubricate
bucket grooves, pivots of
swing-bucket rotors and
threads of the angle rotors
using the lubricant.
Wipe up the moisture in the
chamber with a dry cloth after
use. Keep the lid open when not
in use.
Clean the chamber and
rotor by wiping with a cloth
soaked in mild detergent and
dry completely.
Care of Swing-Out Rotor
Keep the Chamber Dry
Cleaning
Check the rotor, buckets and tubes are
free of scratch and crack before operation.
Check the Accessories for Abnormalities
Load tubes symmetrically into the
rotor with identical amount.
Load Tubes Symmetrically
A-15-24
A-50c-10
A-50-10
S-mw-2
S-15-6
S-50-4
S-100-4
24 x 15 mL / Max. 5k
10 x 50 mL Conical / Max. 5k
10 x 50 mL / Max. 4k
2 x 4 MTP / Max. 4k
6 x 2 x 15 mL / Max. 4.4k
4 x 50 mL / Max. 5k
4 x 100 mL / Max. 4.5k
Rotor
Electrical Safety
Switch off the device after
use. Unplug the power plug
before cleaning or unused for
a long period of time.
Check the wall outlet
is earth grounded,
surge-protected andcan
supply enough power.
Switch Off the Device
Grounded Outlet
Check the Line Voltage
Maintenance
If the centrifuge lid cannot be opened, insert
the provided T-wrench into the emergency
release hole on the left side of the device and
turn it clockwise. (Before activating
emergency open, switch off the device and
wait for the rotor to stop.)
Emergency Lid Open
Swing-out Rotors
Fixed-angle Rotors
Connect the device to voltage
sources which correspond to
the electrical requirements on
the label attached to the device.
900VA
220V 50/60Hz
(110V Optional)
Keep a safety clearance of
30 cm around the centrifuge
during operation.
Empty the Condensate
Keep a Safety Clearance
Empty the condensate
regularly by removing the
drain cap on the left side of
the device
30 cm
30 cm
30 cm
HI-QG-T05R(E)(Rev.1)
Quick Guide
T05R
Cat. No. : T05R