USER MANUAL
T05R
Centrifugal separators are dangerous because they use a high-speed rotating body.
Safety precautions are to prevent personal injury, product damage, and breakdown from possible
dangers during use.
Please observe all safety measures described in this manual.
1.3. Precautions for safety
1. The centrifuge must be installed horizontally on a level surface. If the shaft is operated in an inclined
state, a large vibration may occur or damage to the device may occur.
2. Before connecting the centrifuge to power, check the voltage to be used.
If it is connected with the wrong voltage, it may cause damage to the device and personal injury.
3. Use only the rotor and recommended parts and accessories provided by Hanil Scientific Inc. We are
not responsible for any damage to the device or accidents resulting from the use of non-recommended
parts and accessories.
4. The sample must be used in a centrifugal separator tube, and must be used within the maximum xg
value of the tube.
5. In the case of centrifuging dangerous substances (pathogenic, toxic, radioactive substances, etc.), it is
necessary to sufficiently grasp the physical properties of the substance and take necessary safety
measures.
6. If the centrifuge is contaminated with pathogenic, toxic or radioactive substances, the contaminants
must be thoroughly removed and necessary measures such as ventilation or isolation must be taken.
7. Substances that can generate volatile or explosive vapors cannot be centrifuged.
8. When the rotor comes into contact with cleaning solutions such as strong acids or strong bases or
cesium/silver/salt, it will cause a chemical reaction and corrosion will begin.
9. The rotor chamber must be kept dry at all times before using the centrifuge.
10. Do not drive the rotor above the permitted rotational speed. If the rotor is subjected to a
centrifugal force that exceeds the allowable rotation limit, the rotor will be deformed and damaged.
11. Before centrifugation, the sample must be balanced.
12. It is forbidden to touch or move the rotating rotor.
13. The rotor must be accurately fixed to the rotating shaft, and the rotor used with the rotor lid must
be securely fastened and used. If the lid comes off during rotation, it may cause serious damage
to the product and sample.
14. Do not block the air ventilation for proper air flow which keeps the centrifuge from overheating.
15. Do not put any objects into the openings of the centrifuge.
16. Never use a tool to remove the lid or guard.
17. When requesting repair, the user must remove contaminants in advance.
18. Maintenance must be performed by a technician authorized by Hanil Scientific Inc.
19. For product repair, contact the place of purchase.
20. When operating according to the IEC61010-2-020 standard, the safety distance (30 cm) around the
centrifuge must be observed for smooth instrument operation and the safety of users and
the surrounding environment.
21. Turn off the device switch after using the device.
22. Disconnect and store the power cord from the power outlet before washing the machine or when
not in use for a long time.
Содержание T05R
Страница 1: ...LOW SPEED CENTRIFUGE T05R USER MANUAL...
Страница 7: ...6 A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com USER MANUAL T05R...
Страница 22: ...21 ihanil com A S Inquiry 02 3452 8966 techsupport ihanil com USER MANUAL T05R...
Страница 23: ...22 A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com USER MANUAL T05R...
Страница 25: ...USER MANUAL T05R 7 Rotors and accesories ihanil com 24 A S Inquiry 02 3452 8966 techsupport ihanil com...
Страница 26: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com 25 ihanil com...
Страница 27: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com 26...
Страница 28: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com 27 ihanil com BB100bs...
Страница 29: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com 28 b BBd15cd...
Страница 30: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com 29 ihanil com...
Страница 31: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com 30...
Страница 32: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com 31 ihanil com...
Страница 33: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com 32...
Страница 34: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com MEMO 33 ihanil com...
Страница 35: ...USER MANUAL T05R A S Inquiry 02 3452 8966 techsupport ihanil com ihanil com 34 MEMO...
Страница 36: ......