4. Operation
4.2 Opening and closing the lid
•The lid can only be opened when the centrifuge is switched on.
•Do not reach with your fingers between the housing and lid.
•Close the lid completely before operation.
•
Opening the lid
Press the ( ) lid key.
•
Closing the lid
Put the lid on and press the edge of the lid.
4.3 Loading and removal of the rotor
•Only use rotors, parts, and accessories provided by Hanil Scientific Inc.
•Do not use scatched or cracked rotor in use.
Loading
①
Clean the motor shaft and the chamber.
- When loading the rotor, the temperature should be 10 to 30 degrees.
②
Load the rotor vertically onto the motor shaft.
③
Insert the wrench into the center hole of the rotor
and turn it clockwise to tighten the rotor bolt.
④
Check the rotor to make sure it loaded firmly.
Unloading
①
To remove the rotor,
insert the wrench into the center hole of the rotor
and turn it counterclockwise to loosen the bolt.