4. Operation
4.4 Loading Cytocell accessories
• Only use tubes and accessories provided or approved by Hanil Scientific Inc.
• Always use the same type of tube.
• Tubes should be loaded symmetrically.
• Do not exceed the maximum rated speed of the tube.
• Same volume of sample should be put on tubes.
• Check symmetric loading by balncing tubes with scales.
①
Check the maximum load for each tube.
②
Put cytocell accessories into rotor holes vertically.
③
Cytocell accessories located opposite each other must be the same type and contain
the same quantity.
11 l www.ihanil.com
-Cytocell accessories assembly-
①
Place the rubber plate and filter on the sus holder in this order.
-Loading cytocell accessories-