Hanger ACP MEGAPULSE II Скачать руководство пользователя страница 8

 

MEGAPULSE

®

 II USER MANUAL 

 

 

 

COPYRIGHT © 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED 

 

Keep cables at least several inches away from any objects or material, especially metal or grounded objects.  

 

Do  not  lean  on  or  hold  the  cables  while  the  generator  is  activated.  In  addition  to  the  strong  heating  effect,  a 
deteriorated cable could break down and expose the user to high voltages.  

 

Keep all line cords away from the diathermy unit cables. Do not store or coil line cords where they can come close 
to the cables on an operating diathermy unit.  

 

Before  increasing  generator  output  in  response  to  a  report  of  inadequate  patient  heating,  verify  that  cables  are 
properly routed, evenly spaced, and away from any metal or grounded objects. The heating effect may be "stolen" 
from the patient at a spot where the cables are close to a metal object or each other. Increasing the generator output 
under these circumstances will cause increased heating at that spot.  

 

Use a nonconductive treatment table and a mattress or couch without metal parts (e.g., decorative buttons, springs) 
near the patient, applicators, or cables. Do not use conductive mattresses or mattress covers.  

 

During application of shortwave diathermy, the operator and any other patients should stand at least 2 feet from the 
device.  

 

Do not apply shortwave diathermy over areas with excess adipose tissue.  

 

All equipment and accessories should be kept out of the reach of children or unqualified persons. 

 

Do not apply over areas of hemorrhage or active bleeding.   

 

Avoid applying to the body if wet.  Dry the area thoroughly, as water in the treatment field may lead to uneven 
heating. 

 

All hearing aids should be removed during shortwave diathermy treatment. 

 

Do not turn on the output of the Megapulse

® 

II until the head (Drum) is properly placed above/over the treatment 

area. 

 

Caution  should  be  used  when  applying  thermal  shortwave  over  areas  of  body  which  lack  normal  sensation. 
Absent  or  diminished  sensation  should  be  avoided  or,  if  unavoidable,  treated  with  caution.    Establishment  of 
acceptable intensity levels for desensitized areas may be related to the intensity levels tolerated on normal skin 
in opposite or related body parts. 

Содержание ACP MEGAPULSE II

Страница 1: ...by reference in its entirety This Manual may not be used for any other purpose Any additional copies of the Manual shall be ordered from ACP No changes or modifications shall be made to the Manual wit...

Страница 2: ...Systems Our MEGAPULSE NEUROPROBE OMNISTIM OMNISOUND OMNICYCLE OMNIVR and SYNCHRONY represent the most recent worldwide advances available for therapeutic application of electromedical devices and othe...

Страница 3: ...ts Caution electrical precautions Remote Control connection CE mark of confidence compliant to MDD 93 42 EEC Connection Electrode Cable IP2X Classification against intrusion of water dust accidental c...

Страница 4: ...olation 12 Cleaning Disinfecting of the Megapulse II 12 Cleaning and Low Level Disinfection 12 Intermediate Level Disinfection 13 Use of Barriers 13 MODES OF OPERATION 14 Thermal Effect Variables 14 P...

Страница 5: ...healthcare professional Indications Diathermy is used therapeutically to increase the temperature of tissues It is generally accepted that heat produces the following desirable therapeutic effects Rel...

Страница 6: ...ad or any implanted system that may or may not contain a lead Both the thermal and sub thermal modes of operation pose a risk of tissue destruction o If you are a physician who implants or monitors pa...

Страница 7: ...y over the cranial region as little is known about the electromagnetic effects on the cerebrum to determine if it represents a serious hazard when shortwave diathermy is applied to the head Do not app...

Страница 8: ...le and a mattress or couch without metal parts e g decorative buttons springs near the patient applicators or cables Do not use conductive mattresses or mattress covers During application of shortwave...

Страница 9: ...time and if there is evidence of or if the patient complains of pain or overheating during treatment adjust the output downward until it is tolerated by the patient If the patient continues to complai...

Страница 10: ...ether understood it is well established that this energy causes vibration of ions and molecules within the cells and tissues Depending on the intensity of the high frequency energy applied this vibrat...

Страница 11: ...indicated by the output display LED G Pulse duration switches select the pulse duration in microseconds as displayed in the PULSE DURATION display Pressing the Up Arrow increases the pulse duration to...

Страница 12: ...commonly used pulsed shortwave electrode Drum inductive applicators are generally referred to as monodes Drum inductive applicators are operated in the pulsed mode to avoid overheating of the coils Hi...

Страница 13: ...a depth of 2 inches 5cm Thermal Dosage Pulsed shortwave diathermy provides outputs capable of producing thermal effects Average power output in watts is controlled by the peak power and the duty facto...

Страница 14: ...nfecting of the Megapulse II Modality equipment shall be cleaned disinfected per facility infection control policy ACP recommends the following guidelines Proper cleaning procedures at the end of each...

Страница 15: ...ed to cover any surface that may be touched during a patient treatment to help prevent cross contamination Barrier film is for single use only The film is discarded after each patient treatment The pr...

Страница 16: ...e field applicators Typical values for the Megapulse II drum inductive applicator is 1 5 SAR The higher the SAR value the greater the thermal effect in the tissue III Treatment Time In general treatme...

Страница 17: ...ation following stretch or mobilization making PSWD an effective modality for treatment of scar tissue and contracture Summary of Physiologic Effects 1 Vasodilatation of blood vessels 2 Increase in ce...

Страница 18: ...show 800 PPS If the OUTPUT ON OFF CONTROL 6 switch is pressed with no time set on the display the buzzer will sound and the output will not be energized 3 Audible confirmation of the operation of the...

Страница 19: ...t contact 5 When the drum inductive applicator has been placed in position tighten the hand wheels on the arm to prevent movement Switch treatment ON 6 The protocol LED 3 on the front panel will light...

Страница 20: ...press gently on the drum applicator pin to dislodge the drum applicator from the articulating arm The drum applicator can now be applied to a patient where the articulating arm might have obstructed t...

Страница 21: ...en unit is ON and treatment running drum inductive applicator is not emitting output Patient has impaired sensation over treatment area Use output tester if available to determine if unit has failed o...

Страница 22: ...to May of 2004 the default pulse rate setting is 100 PPS Pulse Duration 400 200 100 65 40 20 s available Pulse Repetition Rate 50 100 200 400 600 800 PPS available Timer The digital treatment timer in...

Страница 23: ...ic Representation of Pulse Trains 20 MICROSECOND Sec PULSES 800 PULSES PER SECOND PPS 65 Sec PULSES 400 PPS 200 Sec PULSES 200 PPS 400 Sec PULSES 100 PPS CONTINUOUS TRAIN OF PULSES PULSES CONTAIN 27 1...

Страница 24: ...R IN WATTS W EFFECT 400 Sec 3 0W 6 0W 12 0W 24 0W 36 0W 48 0W 4o C 200 Sec 1 5W 3 0W 6 0W 12 0W 18 0W 24 0W 2o C 100 Sec 75W 1 5W 3 0W 6 0W 9 0W 12 0W 1o C 65 Sec 49W 1 0W 2 0W 3 9W 5 9W 7 8W SUBTHERM...

Страница 25: ...29000 Megapulse Power Cord Hospital Grade 10ft 72763 Applicator Head Coaxial Cable 89555 Megapulse II User Manual Infection Control Supplies ITEM ITEM NO DESCRIPTION 55536 Super Sani Cloth Wipes Sing...

Страница 26: ...f any of the following have occurred 1 The product contains repairs or replacement parts not furnished by ACP 2 The product is damaged resulting from misuse or negligence 3 The product has been tamper...

Страница 27: ...nty ACP is not responsible for any loss or damage to the Equipment prior to receipt at the ACP Service Center Equipment returned for warranty service must be shipped complete with all accessories exce...

Страница 28: ......

Отзывы: