20
MEGAPULSE
®
II USER MANUAL
COPYRIGHT © 2017, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
TECHNICAL SPECIFICATIONS
Technical Data
GENERAL:
Power Supply:
240, 120 ac 50/60Hz.
Power consumption:
180W max.
Fuses:
Externally Replaceable
2 x T 2A 20mm for 120
Size:
Approx. H: 870 mm (34.25 in) / D: 410 mm (16 in) / W: 410 mm (16 in)
Weight:
39 kg (86 lbs)
Classification:
Class 1: Type BF. (IEC 60601-1: 2001)
Frequency:
27.12 MHz + 20 KHz
Output Power:
150W peak power (+20%) into 50 Ohm load pulsed mode
Output modes:
Pulsed output waveforms are available as selected by the front panel. In pulsed mode the output is square wave
modulated.
Pulse modes presets:
Protocol 1ΔT
- Timer is preset to 20 minutes, pulse rate is set 1
T to 800 PPS and pulse duration to 100 µSec.
Protocol 2ΔT
- Timer is preset to 20 minutes, pulse rate is set 2
T to 800 PPS and pulse duration to 200 µSec.
Protocol 4ΔT
- Timer is preset to 20 minutes, pulse rate is set 4
T to 800 PPS and pulse duration to 400 µSec.
VAR
- Timer is variable from 0 - 30 minutes, but Variable preset at 30 minutes, pulse rate is set to 400PPS, 65
µSec pulse duration. Pulse rate, timer and pulse duration are variable over their full range in this mode.
NOTE:
On Megapulse
®
units manufactured prior to May of 2004, the default pulse rate setting is 100 PPS.
Pulse Duration:
400, 200, 100, 65, 40, 20 µs available
Pulse Repetition Rate:
50, 100, 200, 400, 600, 800 PPS available
Timer:
The digital treatment timer indicates the set time in minutes and seconds prior to the start of treatment and treatment
time remaining during treatment. The maximum setting of the timer is 30 minutes. The timer increments in 1 or 10
minute units.
Timer accuracy:
+/- 1 second at all settings
Electrode:
Hi Q drum inductive applicator
Cooling:
Forced air by integral fan (in drum inductive applicator)
ENVIRONMENTAL CONDITIONS:
For Transport and Storage:
Temperature: -10 to 35oC
Relative Humidity: 5 to 95%
Atmospheric Pressure: 500 to 1060 hPa
Each unit is supplied with a main power cable. Both the cable and the plug on the unit conform to international
safety standards.
All information on supply voltages, model, serial number, and month/year of manufacture is located on the rear
panel. The external fuse ratings and type information is located on the bottom of the unit.
Standards
The Megapulse
®
has been designed to meet the requirements of IEC 60601-1:2001 “Safety of Medical Electrical
Equipment, Part l: General requirements.”
In addition the
Megapulse
®
has been designed to meet the requirements of IEC 60601-2-3:1998 “Specification for
safety for shortwave therapy equipment,” and the requirements of FCC standards 47 CFR Part 18 Subpart C.