background image

15

M A N U A L

System

RoMedic

TM

Symbols

Read user manual

May not be discarded in domestic waste

The product fulfills the demands stated in the MDD 93/42/EEC

Type B, according to the degree of protection against electric shock.

The device is intended for indoor use.

Class lI equipment

Always Read The User Manual

This equipment/system is intended for use by healthcare professionals only. This equipment/system may cause radio 
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such 
as reorienting or relocating the equipment/system or shielding the location.

EMC

English

Содержание SystemRoMedic MiniLift160EE

Страница 1: ...or descriptive matter set out hereon are the confidential a and must not be disclosed loaned copied or used for manufacturing tendering or any oth SCALE PROJECTION Material no MATERIAL SPECIFICATIO A...

Страница 2: ...width adjustment 4 C D B This drawing and any information or descriptive matter set out hereon are the confidential and copyright property of Handicare and must not be disclosed loaned copied or used...

Страница 3: ...function and perform a final inspection see Final inspection Connect cables Hand control cord in HS cable to lift unit on mast in outlet M1 and cable to base width motor in outlet M2 EE To collapse th...

Страница 4: ...y pushing down on the respective pedals for widening and narrowing the base For EE use the buttons on the hand control Test lift function by lifting a person not a user using an approved sling At the...

Страница 5: ...he lift must not be used or immersed in or under water including showering The lift should not be left stored in a high humidity environment The lift must not be charged in a wet room To ensure optima...

Страница 6: ...the battery is fully charged Wall mounted charger 1 Remove the battery pack from the lift and place it in the wall mounted charger 2 Check to ensure that the LED lamp on the front of the charger light...

Страница 7: ...ng 100 50 Battery state 2 Battery needs charging 50 25 Battery state 3 Battery needs charging Less than 25 A tone sounds when a button is pressed in this battery state Battery state 4 The battery need...

Страница 8: ...widen and narrow the base Emergency stop To activate emergency stop Activate the red emergency stop button on the control box Resetting Turn the button in the direction of the arrows until the button...

Страница 9: ...ays supported by the actuator Hand control functions The hand control has four functions up and down lift arm out and in base width Two functions cannot be used at the same time NOTE The hand control...

Страница 10: ...s possible to the user Take care to ensure that the user s shins are resting securely against the leg support and then adjust 2 the leg support so that it does not press against the user s kneecaps ab...

Страница 11: ...ar handles with both hands to avoid being pinched between the slingbar and lift loops 7 6 7 Helpful hint Use a longer loop alternative for short users and for users that cannot be fully raised Use a s...

Страница 12: ...lift makes unusual noises Try to determine the source of the sound Take the lift out of operation and contact your dealer Accessories ThoraxSling polyester XS XXL article no 45500003 009 ThoraxSling...

Страница 13: ...o not use cleaning agents containing phenol or chlorine as this could damage the aluminium and the plastic materials If disinfection is needed 70 ethanol 45 isopropanol or similar should be used Stora...

Страница 14: ...ctric shock The device is intended for indoor use Class lI equipment Always Read The User Manual This equipment system is intended for use by healthcare professionals only This equipment system may ca...

Страница 15: ...per minute Push 4 000N Sound level With load upwards 74 7 dB A downwards 52 6 dB A Material Steel Emergency lowering Mechanical and electrical Castors Front 3 9 100 mm back 3 9 100mm Weight 89 3 lbs 4...