Fortress 1700 DT/TA
49
Possible cause
Action
Action to
be taken
by
•
The battery voltage is too low.
•
Check the battery voltage.
See ‘Battery condition
meter’.
•
Charge the batteries for 8
hours.
•
Check operation of the
battery charger.
Yourself
•
The scooter drives too slowly.
•
Turn the speed control
knob to the right
Yourself
•
Overheated motor.
•
Allow the scooter to cool
down.
Yourself
Sometimes simply switching the scooter off and back on again will correct the
problem.
If you cannot solve the problem by using the list above, contact your dealer.
7.2 Automatic circuit breaker
•
See Fig. 000000180.
The scooter has been equipped with an
automatic circuit breaker (A).
In normal circumstances, the automatic
circuit breaker will not pop out and the
scooter will be ready for use.
When the electric circuit is overloaded or
shorted, the automatic circuit breaker will
switch off the electric system to prevent
damage.
Overload can be caused if the motor is
overheated. If the automatic circuit
breaker switches off the electric circuit of
the scooter, it will pop out.
The automatic circuit breaker is located on the rear cover under the seat.
By pressing the automatic circuit breaker back in again, you can try to restart
the scooter. If the automatic circuit breaker pops out again, wait 15 to 20
minutes to allow the motor to cool off, and then try resetting again.