Periodic inspection - instruction
Check battery and electrical cable.
-Verify that all cables are properly inserted.
Controlbox
-Verify that all functions on the control panel are working
Handcontrol
-Verify that all functions on the handcontrol are working.
Emergency stop
-Verify that the lift does not operate when emergency stop is pressed in.
Emergency lowering
-Verify the function of emergency lowering.
Charging
-Verify that charging is operating. Yellow lamp on the controllbox should light up when charging.
Product lable
- Check the serial number so that the label is there and readable. Note the serial number on the protocol.
Lift arm joint
- Check the joint for cracks or ware.
Max. load test
Mast max. load
-Run the lift with maximum load, all the way up and down, listen for peculiar noises and vibrations.
Base max. load
-Widen the base with maximum load, listen for peculiar noises and vibrations.
Accessories
-Check the accessories to the lift.
Documentation
-Instruction / Manual
To reset the service counters press both lift buttons on the hand control at the same time for 5 seconds.
An audio signal will indicate that the timer has been reset.