Quick Check® SV Series User’s Guide
6 - 35
~PB816xxx100
This command sets the minimum percent of scans on a code which calculate
good quiet zones to determine an acceptable quiet zone analysis. The field xxx
is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of
all fully decoded scans on a code must calculate a good quiet zone, or a failure
condition is set for this parameter.
~PB806xxx000
This command sets the threshold for the ANSI Defects calculation on a code to
set a failure condition. The field xxx is the passing threshold for the calculation.
For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure
condition for this parameter.
~PB802xxx100
This command sets the threshold for the ANSI Decodability calculation on a code
to set a failure condition. The field xxx is the passing threshold for the
calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower
will cause a failure condition for this parameter.
~PB804xxx100
This command sets the threshold for the ANSI Symbol Contrast calculation on a
code to set a failure condition. The field xxx is the passing threshold for the
calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or
lower will cause a failure condition for this parameter.
~PB811xxxyyy
This command sets the threshold for acceptable X dimension ranges. The value
for xxx is the narrowest acceptable X dimension. The value yyy is the widest
acceptable X dimension. Units are in .001 inch (mils). For example:
~PB811010020 will cause a failure for n X dimension less than 10 mils or greater
than 20 mils.
~LAxx
This command sets the passing threshold for the Overall ANSI method Symbol
Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower
calculated for the code analyzed to set a failure condition for this parameter.
~LN##
This command sets the minimum number of codes to be read during a sync
period. ## = the number of codes. For example: ~LN02 causes a No Read
condition to be set if less than 2 bar codes are fully decoded during a sync
interval. (This system is using edge sync mode, therefore a sync interval is the
time between sync inputs.)
~LZ##
This command sets the exact number of codes to be read during a sync period.
## = the number of codes. For example: ~LZ02 causes a No Read condition to
be set if exactly 2 bar codes are not fully decoded during a sync interval.
This command overrides the ~LN## command if the number of codes set does
not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence.
Содержание SV series
Страница 1: ... User s Guide Quick Check SV Series Bar Code Scanner Verifier ...
Страница 8: ...iv ...
Страница 16: ...2 6 Quick Check SV Series User s Guide ...
Страница 22: ...3 6 Quick Check SV Series User s Guide ...
Страница 98: ...8 6 Quick Check SV Series User s Guide ...
Страница 103: ......
Страница 104: ...Hand Held Products Inc 700 Visions Drive P O Box 208 Skaneateles Falls NY 13153 0208 QCOLVSV UG Rev B 5 05 ...