S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
1
HAMEG Instruments GmbH • Industriestraße 6 • D-63533 Mainhausen
Subject to change without notice
Tel.: +49 (0) 6182 800 0 • Fax: +49 (0) 6182 800 100 • E-Mail: [email protected]
Geschäftsführer: Dipl.-Ing. Holger Asmussen • AG Offenbach am Main HRB 41200
Firmware Version: 01.020 and later
English
S C P I P r o g r a m m e r s M a n u a l
H M S S e r i e s
HMO352x, HMO2524
HMO72x ... HMO202x
Firmware Version: 03.00 and later
English
:ACQuire
:CHAN1OFFSet
:DISPlay
:HCOPy
*ESR?
*OPC?
:LOGic:SIZE
:MEASure
:POD1
:SYSTem
*SRE?
S C P I P r o g r a m m e r s M a n u a l
H M O S e r i e s
SCPI Commands HMC8012
2.7.2 STATus:QUEStionable Registers
The commands of the
STATus:QUEStionable
subsystem control the status reporting structures of the
STATus:QUEStionable registers:
See also:
–
chapter 1.6.1, „Structure of a SCPI Status Register“
–
„STATus:QUEStionable Register“
, on page 18
– Diagram on page 20
The following commands are available:
STATus:PRESet ......................................................................................................... 70
STATus:QUEStionable:CONDition? ........................................................................... 70
STATus:QUEStionable:ENABle <Enable_Value> ....................................................... 70
STATus:QUEStionable:ENABle? ................................................................................. 70
STATus:QUEStionable[:EVENt]? ................................................................................ 71
STATus:PRESet
Resets all bits of the STATUS:QUESTIONALBLE and Standard Operation enable register.
Usage:
Event
STATus:QUEStionable:CONDition?
Returns the contents of the CONDition part of the status register to check for questionable instrument or measure-
ment states. Reading the CONDition registers does not delete the contents.
Return values:
<Condition>
Condition bits in decimal representation
Range:
1 to 65535
Increment: 1
Usage:
Query only
STATus:QUEStionable:ENABle
<
Enable_Value
>
Sets the enable mask that allows true conditions in the EVENt part to be reported in the summary bit. If a bit in the
enable part is set to 1 and its associated event bit transitions to true, a positive transition occurs in the summary bit
and is reported to the next higher level.
Parameters:
<Enable_Value>
Bit mask in decimal representation
Range:
1 to 65535
Increment: 1
STATus:QUEStionable:ENABle?
Reads the enable register and returns a decimal value which corresponds to the binary-weighted sum.