Safety

The instrument has been designed and tested in accordance
with VDE 0411 , safety requirements for electrical equipments
for measurement, control and laboratory use. It has left the
factory wholy conform to this standard. Thus it is also in
accordance  with the International standard  IEC Publication
1010-1  and  the European standard EN 61010-1.

The case chassis and all measuring terminals are connected to
the protective earth contact of the appliance inlet. The
instrument operates according to safety Class I (three-conductor
power cord with protective earthing conductor and a plug with
earthing contact).

CAUTION:
The mains/line plug shall only be inserted in a socket
outlet provided with a protective earth contact. The
protective action must not be negated by the use  of an
extension cord without a protective conductor  (or
other means).

The mains/line plug should be inserted before connections are
made to measuring circuits. Whenever it is likely that protection
has been impaired, the instrument shall be made inoperative
and be secured against any unintended operation. The
protection is likely to be impaired if, for example, the instrument

·

shows visible damage.

·

fails to perform the intended measurements.

·

has been subjected to prolonged storage under unfavourable
conditions (e.g. in the open or in moist environments).

·

has been subject to severe transport stress (e.g. in poor
packaging).

In case of leakage coming from a broken «lcd» display and
contact with skin, clean with pure water and soap . When the
metallic case is open or replaced, the instrument has to be
disconnected from the power. If  measurements or calibrations
are necessary with an open instrument, they must be performed
by a qualified technician.

Operating conditions

The instrument has been designed for indoor use. The
permissible ambient temperature range during operation is
+0°C to +40°C, nevertheless  refer to the technical specifications
at +23°C 2°C. It may occasionally be subjected to temperatures
-10°C without degrading its safety. The permissible ambient
temperature range for storage or transportation is -40°C to
+60°C.

The maximum operating altitude is up to 2200m. The maximum
relative humidity is up to 80%.

If condensed water exists in the instrument it should be
acclimatized before switching on . In some cases (e.g. extremely
cold) two hours should be allowed before the instrument is put
into operation. The instrument should be kept in a clean and
dry room and must not be operated in explosive, corrosive,
dusty, or moist environments. The instrument can operate in
any position, but the convection cooling must not be impaired
and the ventilation holes must not be  covered.

Warranty

HAMEG warrants to its customers that the products it
manufactures and sells will be free from defects in materials
and workmanship for a period of 2 years.

This warranty shall not apply to any defect, failure or damage
caused by improper use or inadequate maintenance and
care.This warranty is limited to repair and if necessary to
replace the instrument according to the decision of HAMEG
only.

HAMEG shall not be obliged to provide service under this
warranty to repair damage resulting from attempts by personnel
other than HAMEG represantatives to install, repair, service or
modify these products. In order to obtain service under this
warranty, customers must contact and notify the distributor
who has sold the product.

Each instrument is subjected to a quality test with 24 hours
burn-in before leaving the production . Practically all early
failures are detected by this method. In the case of shipments
by post, rail or carrier it is recommended that the original
packing is carefully preserved. Transport damages and damage
due to negligence are not covered by the guarantee. In the
case of a complaint, a label should be attached to the housing
of the instrument which describes briefly the faults observed.
If at the same time the name and telephone number (dialing
code and telephone or direct number or department designation)
is stated for possible queries , this helps towards speeding up
the processing of guarantee claims.

Power requirement

The instrument is designed for a power source of  110V/220V
50Hz. The frequency and the voltage can fluctuate to  10%. An
AC power cord is supplied with the instrument. Connect it to
the power receptacle on the rear panel. Before connecting the
instrument to the power source, always make certain that the
line voltage switches on the rear panel are set to the position
corresponding to the voltage of the AC power source. The
instrument is protected by a slo-blow fuse installed in the fuse
holder.

General informations

Содержание HM8134-2

Страница 1: ...Instruments MANUAL HANDBUCH MANUEL Programmable Synthesizer HM8134 2 ENGLISH...

Страница 2: ...LEER...

Страница 3: ...d HO90 13 RS232 interface option HO89 13 IEEE 488 interface option HO88 13 PREV key 13 ON OFF key 13 RCL STO keys 13 REMOTE OPERATION 13 Interfaces 13 Commands supported 14 General 14 Description 14 C...

Страница 4: ...d on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residential commercial and light indus...

Страница 5: ...litudechanges samerangefrequency and without modulation is another outstanding chara cteristic of this instrument The instrument parameters are all clearly displayed on two lines of 20 characters each...

Страница 6: ...1 OFF 0 ON General Interfaces options bus IEEE 488 HO88 or RS232 HO89 IEEE 488 functions T6 L4 SH1 AH1 RL1 DC1 DT0 and R0 HO80 Set up memory locations 10 Dimensions 285 X 75 X 365 W X H XD Weight appr...

Страница 7: ...water exists in the instrument it should be acclimatizedbeforeswitchingon Insomecases e g extremely cold two hours should be allowed before the instrument is put into operation The instrument should...

Страница 8: ...MENU Input key in the configuration menu 8 DISPLAY Two lines of 20 characters each on a backlight LCD 9 FUNCTIONS Functions keys and leds 10 DIGITAL ROTARY CONTROL 11 NUMERIC KEYPAD Input parameters w...

Страница 9: ...r to be back in the main display unselecting the current function 9 or cancelling the numeric keypad input Setting parameters Once a parameter is selected for exemple FREQ with a function key 9 a new...

Страница 10: ...rom the data keypad 11 or modified by the digital rotary 10 or one of the four context sensitive keys 5 For more details refer to the paragraph setting parameters The level range is 127dBm to 13dBm wi...

Страница 11: ...here of 20Hz CH1 modulation narrow pulse of 200 s wide pulse of 1 8ms synchronisation bit of 14ms CH2 zoom of CH1 CH1 demodulated signal DC coupled NUM position CH2 demodulated signal AC coupled NUM p...

Страница 12: ...s deviation can be entered from the data keypad 11 or modified by the digital rotary 10 or one of the four context sensitive keys 5 The deviation range may be change by step of 100Hz from 200Hz to 150...

Страница 13: ...modulation The GATE modulation is made by sampling the output carrier with logic signal GATE and is caracterised by on off Ratio Rise Fall time Delay time The GATE signal level TTL is applied at the...

Страница 14: ...keys Ref 5 The phaselock loop of the reference is tested Locked Unlocked Increased stability for the HM8134 may also be obtained from an external oscillator The external reference frequency must be c...

Страница 15: ...ter pressing the STO key 6 we get The current configuration can be stored by pressing a numeric key from 0 to 9 After pressing the RCL key 4 we get A configuration can be recalled by pressing a numeri...

Страница 16: ...alisation RST RST RST RST RST idem key ESC at power on time except beep display com and memory config 0 9 which are not modified General commands IDN identification SAV x store current state x from 0...

Страница 17: ...ou to set the validation level of the GATE modulation the NORMal parameter is for a high level validation and INVert is for a low level validation The line 4 asks the enabling level state The instrume...

Страница 18: ...eters above FM STATe 0 OFF 1 ON 9 FM STATe 10 The line 9 allows you to turn the FM on or off The parameters 1 or ON turn it on and 0 or OFF turn the modulation off if present The line 10 asks the curr...

Страница 19: ...the same as the options in the external FM menu refer to the corresponding paragraph The line 14 asks the current mode to the instrument Itsendsbackthesamestringsasthecorresponding command parameters...

Страница 20: ...Hz 400kHz 63 FM deviation error must be in the range 1kHz 200kHz 64 FM deviation error must be in the range 200Hz 150kHz 70 AM frequency modulation error must be in the range 10Hz 20kHz 71 AM frequenc...

Страница 21: ...21 Subject to change without notice Flowchart Menu Function selection...

Страница 22: ...22 Subject to change without notice Flowchart Menu Step Control...

Страница 23: ...23 Subject to change without notice Flowchart Menu Amplitude Modulation Control...

Страница 24: ...24 Subject to change without notice Flowchart Menu Phase Modulation Control...

Страница 25: ...25 Subject to change without notice Flowchart Menu Frequency Modulation Control...

Страница 26: ...26 Subject to change without notice Flowchart Menu Gate Control...

Страница 27: ...27 Subject to change without notice Flowchart Menu Main Menu Control...

Страница 28: ...28 Subject to change without notice Referency loop CREF Synoptic Principal loop PLL1 134...

Страница 29: ...29 Subject to change without notice Secondary loop PLL2 134 Synoptic Transposition TRA 134...

Страница 30: ...30 Subject to change without notice Frequency synthesize DDS134 Level lock loop AMPL134 Synoptic...

Страница 31: ...31 Subject to change without notice Programmable attenuator ATP134 Synoptic...

Страница 32: ...58 3 76 0 83 10 76 0 09 1 20 0 34 2 03 0 59 3 88 0 84 11 50 0 10 1 22 0 35 2 08 0 60 4 00 0 85 12 33 0 11 1 25 0 36 2 13 0 61 4 13 0 86 13 29 0 12 1 27 0 37 2 17 0 62 4 26 0 87 14 38 0 13 1 30 0 38 2...

Страница 33: ...8 18 4 1 860 14 4 1 174 10 4 0 740 6 4 0 467 2 4 0 295 18 3 1 839 14 3 1 160 10 3 0 732 6 3 0 462 2 3 0 291 18 2 1 818 14 2 1 147 10 2 0 724 6 2 0 457 2 2 0 288 18 1 1 797 14 1 1 134 10 1 0 715 6 1 0...

Страница 34: ...4 27 542 10 4 10 965 6 4 4 365 2 4 1 738 18 3 67 608 14 3 26 915 10 3 10 715 6 3 4 266 2 3 1 698 18 2 66 069 14 2 26 303 10 2 10 471 6 2 4 169 2 2 1 660 18 1 64 565 14 1 25 704 10 1 10 233 6 1 4 074...

Страница 35: ...5 7 6 095 20 3 10 351 2 0 1 259 6 6 2 138 11 2 3 631 15 8 6 166 20 4 10 471 2 1 1 274 6 7 2 163 11 3 3 673 15 9 6 237 20 5 10 593 2 2 1 288 6 8 2 188 11 4 3 715 16 0 6 310 20 6 10 715 2 3 1 303 6 9 2...

Страница 36: ...2 51 48 251 68 2 51 108 251 128 2 51 9 2 82 49 282 69 2 82 109 282 129 2 82 10 3 16 50 316 70 3 16 110 316 11 3 55 51 355 71 3 55 111 355 12 3 98 52 398 72 3 98 112 398 13 4 47 53 447 73 4 47 113 447...

Страница 37: ...37 Subject to change without notice...

Страница 38: ...38 Subject to change without notice...

Страница 39: ...LEER...

Страница 40: ...6182 800 0 Telefax 49 0 6182 800 100 E mail sales hameg de service hameg de Printed in Germany Oscilloscopes Multimeters Counters Frequency Synthesizers Generators R and LC Meters Spectrum Analyzers P...

Отзывы: