5
Subject to change without notice
S p e c i f i c a t i o n s
150 MHz Analog Oszilloscope HM1500-2
Valid at 23 °C after a 30 minute warm-up period
Vertical Deflection
Channels:
2
Operating Modes:
CH 1 or CH 2 separate,
DUAL (CH 1 and CH 2 alternate or chopped),
Addition
X in XY-Mode:
CH 1
Invert:
CH 1, CH 2
Bandwidth (-3dB):
2 x 0 - 150 MHz
Rise time:
‹ 2.3 ns
Overshoot:
max. 1 %
Bandwith limiting
(selectable)
:
about 20 MHz (5 mV/ cm - 20 V/ cm)
Deflection Coefficients(CH1,2):
14 calibrated steps
1 mV – 2 mV /cm:
± 5 % (0 - 10 MHz (-3 dB))
5 mV – 20 V /cm:
± 3 % (1-2-5 sequence)
variable
(uncalibrated)
› 2.5 :1 to › 50 V/cm
Inputs CH 1, 2:
Input Impedance:
1 MΩ II 15 pF
Coupling:
DC, AC, GND (ground)
Max. Input Voltage:
400 V (DC + peak AC)
Y Delay Line:
70 ns
Measuring Circuits:
Measuring Category I
Auxiliary input:
Function (selectable):
Extern Trigger, Z (unblank)
Coupling:
AC, DC
Max. input voltage:
100 V (DC + peak AC)
Triggering
Automatic (Peak to Peak):
Min. signal height:
5 mm
Frequency range:
10 Hz - 250 MHz
Level control range:
from Peak- to Peak+
Normal (without peak)
Min. signal height:
5 mm
Frequency range:
0 - 250 MHz
Level control range:
–10 cm to +10 cm
Operating modes:
Slope/Video
Slope:
positive, negative, both
Sources:
CH 1, CH 2, alt. CH 1/2 (
≥
8 mm), Line, Ext.
Coupling:
AC:
10 Hz-250 MHz
DC:
0 -250 MHz
HF:
30 kHz–250 MHz
LF:
0 -5 kHz
Noise Rej. switchable
Video:
pos./neg. Sync. Impulse
Standards:
525 Line/60 Hz Systems
625 Line/50 Hz Systems
Field:
even/odd/both
Line:
all/line number selectable
Source:
CH 1, CH 2, Ext.
Indicator for trigger action:
LED
External Trigger via:
Auxiliary Input (0.3 V
pp
, 150 MHz)
Coupling:
AC, DC
Max. input voltage:
100 V (DC +peak AC)
2nd Trigger
Min. signal height:
5 mm
Frequency range:
0 – 250 MHz
Coupling:
DC
Level control range:
-10 cm to +10 cm
Horizontal Deflection
Operating modes:
A, ALT (alternating A/B), B
Time base A:
0.5 s/cm - 50 ns/cm (1-2-5 sequence)
Time base B:
20 ms/cm - 50 ns/cm (1-2-5 sequence)
Accuracy A and B:
± 3 %
X Magnification x10:
to 5 ns/cm
Accuracy X x10:
± 5 %
Variable time base A/B:
cont. 1:2.5
Hold Off time:
var. 1:10 LED-Indication
Bandwidth X-Amplifier:
0 – 3 MHz (-3 dB)
XY phase shift ‹ 3°:
‹ 220 kHz
Operation/Measuring/Interfaces
Operation:
Autoset, Menu and help functions (multilingual)
Save/Recall (instrument parameter settings):
9
Signal display:
max. 4 traces
CH 1, 2 (Time Base A) in combination with
CH 1, 2 (Time Base B)
Frequency counter:
6 digit resolution:
›1 MHz – 250 MHz
5 digit resolution:
0.5 Hz – 1 MHz
Accuracy
50 ppm
Auto Measurements:
Frequency, Period, Vdc, Vpp, Vp+, Vp-
Cursor Measurements:
Δ
t, 1/
Δ
t (f), t
r
,
Δ
V, V to GND, ratio X, ratio Y
Resolution Readout/Cursor:
1000 x 2000 Pts
Interfaces (plug-in):
RS-232 (HO710),
Optional:
Dual-Interface USB/RS232, IEEE-488 (GBIP)
Dual-Interface Ethernet/USB
Display
CRT:
D14-375GH
Display area (with graticule):
8 cm x 10 cm
Acceleration voltage:
ca. 14 kV
General Information
Component tester:
Test voltage:
approx. 7 V
rms
(open circuit), approx. 50 Hz
Test current:
max. 7 mA
rms
(short circuit)
Reference Potential:
Ground (safety earth)
Probe ADJ Output:
1 kHz/1 MHz square wave signal
0.2 V
pp
(tr ‹ 4 ns)
Trace rotation:
electronic
Line voltage:
105 – 253 V, 50/60 Hz ± 10 %, CAT II
Power consumption:
37 Watt at 230 V, 50 Hz
Protective system:
Safety class I (EN61010-1)
Weight:
5.6 kg
Cabinet (W x H x D):
285 x 125 x 380 mm
Ambient temperature:
0° C ...+40° C
Accessories supplied:
Line cord, Operating manual, 2 Probes 10:1 with
attenuation ID
Optional accessories:
HO720 Dual-Interface RS-232/USB
HO730 Dual-Interface Ethernet/USB
HO740 Interface IEEE-488 (GPIB)
HZ70 Opto-Interface (with optical fiber cable)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com