16 

 

Measuring Flatness 

 

To measure flatness, a horizontal, continuously rotating laser plane is “bucked in” or adjusted so that it is 

parallel to three reference points on a table, set of ways, or a surface. 

 

1.

 

Place all the targets on one reference point and adjust them up or down so they detect the laser 

plane.  

 

2.

 

Zero the targets. 

 

3.

 

Reposition the targets so that one target sits on each reference point.  

 

4.

 

Using the PITCH and ROLL adjustments, adjust the laser scan plane until all three targets 

produce the same value or zero, thus making the laser parallel to the reference points.  

This may also be accomplished by using one target, zeroed on the closest reference point to the laser, 

and moving it back and forth from the reference points until it produces a reading of zero at all three 

points. 

 

5.

 

Move the target to user-specified points on the surface. 

The resulting reading is a measure of the deviation from the reference point, helping to produce a 

flatness profile. The measurement will show either a plus (+) or a minus (-) sign. A plus reading 

indicates that the target is 

higher

 than the reference points and a minus reading means the target is 

lower

 than the reference points. 

 

Measuring Squareness 

 

After bucking in the laser to the five reference points described in 

Measuring Straightness and Measuring Flatness, (to determine the 

straightness and flatness of the machine's axes) measuring 

squareness is a simple process. To 

truly

  measure squareness, one 

must compare the least-squares, best-fit line of the one axis to the 

other axis. If this is not done, bad reference points or severely worn 

ways might produce what looks like a squareness error, but in fact 

is not. To facilitate this type of analysis, our software programs may 

be used to automatically calculate the best-fit line.  

 

To measure Y-to-Z squareness: 

 

1.

 

Lower the column/spindle to its lowest Z position and 

position a target horizontally to pick up the vertical laser 

plane that is perpendicular to the X-axis (parallel to the 

Y-axis). 

2.

 

Zero the target and traverse (raise) the column along its axis.  

The data produces a measurement of both the straightness of the Z-axis and the squareness of the Y-

axis to the Z-axis. 

 

To measure Z-to-X squareness: 

 

1.

 

Position and zero the target to detect the vertical laser plane that is parallel to the X-axis.  

2.

 

Traverse the column upward.  

The resulting data is a measure of the Z flatness and Z-to-X squareness. 

 

To measure X-to-Y squareness: 

 

1.

 

Position and zero the target to detect the vertical laser plane that is parallel to the X-axis. 

2.

 

Traverse the table or column (whichever is moveable) along the Y-axis. 

The result is a measure of both Y straightness and X-to-Y squareness. 

 

Figure 11 – 

The X, Y and Z Axes 

Содержание Dual Scan L-732

Страница 1: ......

Страница 2: ...ing the alignment measurements For best results do not start taking measurements until the recommended warm time has passed It is always a good metrology practice to periodically check your reference...

Страница 3: ...ticles of equipment manufactured by other companies and limited by a warranty extending for less than one year In such an event the more limited warranty applies to said instrument or article of equip...

Страница 4: ...al Pitch Axis 13 Checking the Levels for Accuracy 13 Measurement Overview 14 Selecting Reference Points on Machine Tools 14 Selecting Reference Points on Process Mills 14 Measurement Summary 15 Measur...

Страница 5: ...sfer line wing bases Injection molding machines and presses Aircraft assembly seat track alignment setting water butt and station planes wing to body and body to body join alignment etc Roll paralleli...

Страница 6: ...and targets fit into a small portable shipping case Diode laser 2 times more stable than HeNe based laser systems Battery or AC powered Laser planes flat to arc seconds 00003 in ft or 0 0025mm m in 18...

Страница 7: ...to body and body to body join alignment etc Saw mills Vertical press alignment Measuring and aligning The flatness of almost any horizontal or vertical surface squares frames ways flanges circles etc...

Страница 8: ...p times Applications for alignments with tolerances of 00002 in ft or 0 0017 mm m or higher Roll parallelism in paper mills printing presses and film lines Leveling almost any surface squares frames w...

Страница 9: ...m stability Applications for alignments with tolerances of 00015 in ft or 0 01 mm m or greater Normal or blown film lines roll alignment Rubber roll alignment Leveling almost any surface squares frame...

Страница 10: ...ine adjustments see Figure 1 Number of laser planes the L 743 and the L 733 have three laser planes and the L 742 and L 732 have two laser planes that can be factory configured for two vertical planes...

Страница 11: ...parallel to your references However doing this may result in some laser drift Accordingly if you use the laser prior to the warm up period it is recommended that you go back and check your references...

Страница 12: ...on speed control This control slows the scanner spin until you can see the laser beam pass over the target the farther away the target is located the slower the turret must spin Figure 3 shows control...

Страница 13: ...ure calibrates only one level vial at a time and must be repeated for the other axis Note 2 It is very important to warm up the laser for at least 30 minutes before starting this procedure It is also...

Страница 14: ...alibration of the Level Vials Roll Axis Note It is very important to warm up the laser for at least 30 minutes before starting this procedure It is also very important to level both the Pitch and Roll...

Страница 15: ...tance between the points The result is the calibration of the level vial in units of in ft or mm m To be within specifications the calibration result should be as follows Split Prism Vial Assembly 000...

Страница 16: ...ulating the Calibration of the Levels Vials Pitch Axis 1 Level the laser Rotate the entire laser unit 90 to calibrate the Pitch Axis level vial Using the adjustment knobs level the laser so that both...

Страница 17: ...Pitch Axis level you are adjusting Using the wrench provided adjust the level assembly until the bubble is centered in the window for the Standard Level vials or the two halves of the bubbles line up...

Страница 18: ...rence points carefully Poor reference points like those on the heavily worn portion of a machine bed where all the work is performed may mean extra time to bring the machine back into tolerance In oth...

Страница 19: ...d one vertical laser plane are needed 1 Mount a target horizontally at the closest reference point to the laser and adjust the target so that it detects the laser 2 Zero the target and move it to the...

Страница 20: ...ts described in Measuring Straightness and Measuring Flatness to determine the straightness and flatness of the machine s axes measuring squareness is a simple process To truly measure squareness one...

Страница 21: ...e error from the laser not being parallel to the surface Level to Earth Measurements The leveling of machine tools surface plates and different types of bases can be performed using scanning lasers as...

Страница 22: ...ny three points on a surface may be used however Hamar Laser recommends the setup illustrated in Figure 12 Figure 12 Three Point Buck in Using One Target recommended setup When performing this procedu...

Страница 23: ...ing the Three Point Buck In 1 Center the target in the Near Position With the target in the Near Position zero the target 2 Move the target to Far Position A and tilt the laser beam until the readout...

Страница 24: ...he laser beam must be adjusted to be parallel to or coincident with an edge or centerline The process of making that adjustment is called bucking in This section describes two types of buck in methods...

Страница 25: ...resents the distance from the laser to the first target L2 repre sents the distance be tween the two targets If L1 is less than one tenth of L2 the normal buck in procedure is used If L1 is greater th...

Страница 26: ...ding is to L2 Stated mathematically the ratio is Set L1 Far L2 If L1 L2 and the far reading are known the set can then be determined by the following formula Set Far reading L1 L2 Note This is a simpl...

Страница 27: ...two targets When using normal buck in the rule is Zero Near Point Far Buck in the laser beam by zeroing it on the near target and then pointing the laser beam using the appropriate adjustment knobs t...

Страница 28: ...eck A C power source Use second A C power supply if available 5 Noisy target A 1519 1520 readings on PDA NOTE On average the user can expect 00002 in ft to 00005 in ft 0 0025 mm m 0 004 mm m of noise...

Страница 29: ...low or high on target window it should be near the window It is possible for the laser to activate the auto wake up feature of target but not enough of the beam is hitting the PSD target sensor to get...

Страница 30: ...ime Error in software Do not unplug the A 910 from the USB port while still using the program Report to HLI the exact key strokes that created the Runtime Error 12 Software crashes upon loading Ensure...

Страница 31: ...27 Appendix B Equipment Drawings L 733 Precision Geometry Laser...

Страница 32: ...28 L 743 Ultra Precision Geometry Laser...

Страница 33: ...29 The L 732 Dual Scan Precision Roll Alignment Laser...

Страница 34: ...30 The L 742 Dual Scan Ultra Precision Roll Alignment Laser...

Страница 35: ...ted This makes is very likely that the error in measurement at the point where the target is zeroed is nearly identical to the error in each subsequent measurement because the difference in sensor are...

Страница 36: ...ea is 1 micron or less Approx 4 mm of measuring area each vertical line on the graph equals 1 mm Total Error 3 3 microns 2 microns 2 microns 0 000 Each green vertical line equals 1 mm of measuring are...

Страница 37: ...the type intended for nail polish removal NOTE When cleaning precision optics even with the best quality optical cleaning tissue use gentle pressure to avoid scratching the surface or damaging the opt...

Отзывы: