(4)  External synchronous mode 3 (“EXT.LEVEL1” mode) 

Data is acquired in synchronization with an external trigger signal pulse supplied 
from the CCD. Since this trigger signal pulse is generated in synchronization with 
the X-ray exposure time this mode is also called the "Auto Trigger” mode. Even 
in this mode the CCD performs a dummy scan until the input of an external 
trigger pulse. Upon the input of an external trigger pulse, the CCD carries out 
integration in the same intervals and then outputs the data. The trigger signal 
polarity is negative. 

 

(5)  External synchronous mode 4 (“EXT.LEVEL2” mode) 

This mode is basically the same as the “EXT.LEVEL1” mode. Data is acquired in 
synchronization with an external trigger signal input via the BNC connector of the 
C9266-03/C9266-04.The input signal level is H-CMOS compatible. The trigger 
signal can be either a positive-going or negative-going pulse, selectable by the 
software. 
 

                                    Because there is not a BNC connector, this mode is not usable in the C9266-04. 

 

5)  Exposure time adjustment 

Use to adjust the C9266-03/C9266-04 hardware exposure time for CCD in steps from 
1 to 65,535 mSec. But in the case of "EXT.EDGE1" data acquisition mode, it ranges 
from 41 to 65,535 mSec. If data acquisition mode is Internal mode ("INT" mode), 
External synchronous mode 1 ("EXT.EDGE1" mode) or External synchronous mode 
2 ("EXT.EDGE2" mode), this setting value becomes effective. This setting value 
shows the "Exposure Time" of "5.2 Data acquisition timing charts". 

 

6)  Operating mode setting 

The C9266-03/C9266-04 works in the following 4 operating modes. Operating modes 
(1) to (3) can be selected by the software, but "Data Transfer" mode (4) is 
automatically called up inside the C9266-03/C9266-04 module. The LED display on 
the C9266-03/C9266-04 lets you know which operating mode is currently selected. 

 

(1) “Suspend 

mode” 

This is normal USB operating mode and you can safely disconnect the 
C9266-03/C9266-04 module from the PC. 

 

(2) “Idle/Sleep 

mode” 

Power is being supplied to the C9266-03/C9266-04 module but the CCD is idle 
(not operating). You can connect or disconnect the CCD from the 
C9266-03/C9266-04. 

 

(3) “Standby 

mode” 

The C9266-03/C9266-04 is ready to acquire data in this mode. The CCD is 
repeating a so-called "dummy scan" by sweeping the dark current. 

 

(4)  “Data Transfer” mode 

The CCD acquires an X-ray image and transfers the image data to the PC. 

 

Содержание C9266-03

Страница 1: ...proper handling and operation that does not comply with the instructions in this manual may result in serious accident including injury or death Keep this instruction manual handy so that you can easi...

Страница 2: ...semble or modify any part of this product Malfunctions might otherwise occur b Be careful not to drop bump or apply excessive impacts to this product Drop impacts or bumps may damage the product c Nev...

Страница 3: ...and documentation are reserved to us 4 Reproduction Users can reproduce this software for backup purposes in compliance with the terms of any provision of this license 5 Prohibition clause The user c...

Страница 4: ...Hardware setup 4 2 3 Installing the driver 5 3 Operation 21 4 Functions 22 5 Specifications 24 5 1 Specifications 24 5 2 Data acquisition timing charts 25 5 3 Pin arrangement 28 5 4 Dimensional outlin...

Страница 5: ...with all these functions the C9266 03 C9266 04 still offers compact size light weight and easy use The C9266 03 C9266 04 has five data acquisition modes One is an internal sync mode Internal mode tha...

Страница 6: ...are fed to the CCD through this connector Analog video signals from the CCD proportional to X ray intensity levels and synchronized timing signals H CMOS logic level signals are input to this connecto...

Страница 7: ...C9266 04 5 Power switch The power switch used to turn the C9266 03 on or off Select how power is supplied by using the switch position as shown in the table below Switch position Description Left Pow...

Страница 8: ...re for use with the C9266 03 C9266 04 CCD Data Acquisition Unit C9266 03 C9266 04 Front Panel C9266 03 C9266 04 Rear Panel DC 5 0 V CCD S10810 S10811 USB 2 0 DC Power Supply or AC Adapter Pulse Genera...

Страница 9: ...he Camera module to the USB connector of the PC and ii Connecting the module or the Camera module to an external power source if any 2 3 1 Compliant operating systems This driver is compliant with the...

Страница 10: ...ating from Microsoft s home page Hereafter abbreviated as HP After the Service Pack is installed carry out Windows Update Regarding the procedure and details for Service Pack please refer to Microsoft...

Страница 11: ...at HAMAMATSU xxxxx or HPK xxxxx xxxxx Product number is displayed click the Search for a suitable driver for my device recommend radio button to select it and then click the Next button 5 In the follo...

Страница 12: ...er the HAMAMATSU xxxxx or HPK xxxxx xxxxx Product number driver has been identified and UsbCamIF inf has been confirmed in the driver location click the Next button 7 After the required files are copi...

Страница 13: ...installed right click on My Computer and select Properties The System Properties window will be displayed In this window click the Hardware tab and then click the Device Manager button Check whether...

Страница 14: ...Switch on the PC When the USB camera module is connected to an external power source switch it on after the PC has been turned on 3 Check that HAMAMATSU xxxxx or HPK xxxxx xxxxx Product number is dis...

Страница 15: ...e best driver in these locations radio button and the Search removable media floppy CD ROM check box to select it and click the Next button 5 After the required files are copied confirm that the follo...

Страница 16: ...installed right click on My Computer and select Properties The System Properties window will be displayed In this window click the Hardware tab and then click the Device Manager button Check whether...

Страница 17: ...e PC When the USB camera module is connected to an external power source switch it on after the PC has been turned on 2 Insert the CD ROM that is provided Right click on My Computer and select Propert...

Страница 18: ...cquisition Module C9266 03 C9266 04 14 Right click on HAMAMATSU USB Camera Module and select Properties Click the Driver tab and select Update Driver 3 Click the Next button in the following window Cl...

Страница 19: ...known drivers for this device so that I can choose a specific driver radio button to select it and then click the Next button 5 In the following window select HAMAMATSU USB Camera Module from the Mod...

Страница 20: ...66 03 C9266 04 16 6 Confirm that the following window is displayed and then click the Next button 7 After the following window is displayed click Browse and specify the path as shown below and click O...

Страница 21: ...the Finish button The driver is now upgraded 9 After the driver is installed right click on My Computers and select Properties The System Properties window will be displayed In this window click the...

Страница 22: ...is connected to an external power source switch it on after the PC has been turned on 2 Right click on My Computer and select Properties The System Properties window will be displayed In this window c...

Страница 23: ...erting the CD ROM that is provided click the Install the software automatically Recommended radio button to select it and then select the Next button 4 After the required files are copied confirm that...

Страница 24: ...installed right click on My Computer and select Properties The System Properties window will be displayed In this window click the Hardware tab and then click the Device Manager button Check whether...

Страница 25: ...4 comes with dedicated application software C9266DCamAPL Use this software for control of C9266 03 C9266 04 and data acquisition For information on how to use the C9266DCamAPL software refer to the se...

Страница 26: ...a trigger signal from the application software The other mode is External Trigger mode which acquires data with a trigger signal from an external device The External Trigger mode is further subdivided...

Страница 27: ...66 03 C9266 04 hardware exposure time for CCD in steps from 1 to 65 535 mSec But in the case of EXT EDGE1 data acquisition mode it ranges from 41 to 65 535 mSec If data acquisition mode is Internal mo...

Страница 28: ...Min Scanning rate 1MHz Frame readout time 1 62 sec 2 18 sec Data transfer time 1 62 sec 2 17 sec Total cycle time 1 62 sec 2 18 sec AD conversion resolution 12 bits 4095ADU Conversion gain 48 8e ADU R...

Страница 29: ...r Time Data Acqusition Time Exposure Time Exposure Time Trigger In Softw are Exposure Signal P1V P2V P1H P2H AD Conversion Data Transfer USB2 0 Data Transfer USB1 1 Data Transfer Time Data Transfer Ti...

Страница 30: ...a Transfer Time Data Acquisition Time Dummy Scan Dummy Scan Exposure Time Exposure Time External Trigger In fromBNC Exposure Signal P1V P2V P1H P2H AD Conversion Data Transfer USB2 0 Data Transfer USB...

Страница 31: ...NC connector this mode is not usable in the C9266 04 Data Transfer Time Data Transfer Time Data Transfer Time Data Transfer Time Data Acqusition Time Dummy Scan Dummy Scan Exposure Time Exposure Time...

Страница 32: ...3V 20 P2H 0V 12V 3 SG 0V 12V 21 NC 4 TRIGGER 8V 13V 22 P1H 0V 12V 5 RG 0V 12V 23 NC 6 NC 24 GND 8V 7 OG 11V 25 NC 8 NC 26 RD 20V 9 RD 20V 27 NC 10 NC 28 OD 23V 11 OD 23V 29 NC 12 NC 30 GND 8V 13 OUT 3...

Страница 33: ...CCD Data Acquisition Module C9266 03 C9266 04 29 5 4 Dimensional outlines 1 C9266 03 dimensional outlines...

Страница 34: ...CCD Data Acquisition Module C9266 03 C9266 04 30 2 C9266 04 dimensional outlines...

Страница 35: ...the product or you suspect possible trouble please check it again This is important in order to clarify the failure symptoms and avoid making faulty judgments 5 If a failure is found in the product or...

Страница 36: ...CCD Data Acquisition Module C9266 03 C9266 04 32 7 Document History Date Document Revision Contens 07 Aug 2009 1 00 First edition...

Страница 37: ...u Hamamatsu City Shizuoka Pref 435 8558 Japan Phone 81 53 434 3311 Fax 81 53 434 5184 Document Number K46 B60041 Document Revision 1 00 01 Jun 2009 Information in this document is subject to change wi...

Отзывы: