Instruction Manual KAL 100/200
8
11 12 13
Fig. 3
Rear
10
USB port
11
Inlet connector for non-heating apparatus
12
Micro fuse, 315 mAT, 5 x 20 mm
13
Primary on/off switch
14
Power 24V/125mA, galvanic separation
15
Ground for 24V/125mA
16
Input port for current measurement 0…20 mA,
input resistance 240 Ohm
17
Ground connector for voltage and current input
port
18
Voltage input 0…10V, Ri = approx. 40kOhm
4.2.1 Power input, primary on/off switch, micro fuse
The KAL 200 calibration device is designed at the factory to
accommodate a supply voltage of 230 VAC/50-60 Hz (115 VAC/50-60
Hz is optional). Voltage fluctuations of +6% to -15% are permissible.
The supply voltage connector (11) is located on the rear of the
instrument (inlet connector for non-heating app ground wire).
The instrument’s micro fuse (12) is located above this connector. Value:
315 mAT; dimensions: 5 x 20 mm. Located above the micro fuse is the
double-pole, primary on/off switch (13), which separates the KAL 200
from the supply voltage.
WARNING! Risk of Electrical shock! Failure to unplug the power supply cord
before replacing the fuse may result in fatal injuries!
10
14
15
16
17
18
11
13
12