READ AND SAVE THESE INSTRUCTIONS

Operation Manual 

110V | 220V

STD

1911

®

Содержание HaloFogger STD 1911

Страница 1: ...READ AND SAVE THESE INSTRUCTIONS Operation Manual 110V 220V STD 1911...

Страница 2: ...treatment process Reservoir Door Funnel Port Status Indicator Light Walk Around Tour of the HaloFogger Status Indicator Light Displays the operational status of the Halo device Solid Green Indicates t...

Страница 3: ...s when pouring disinfectant fluid into the device Avoid splashing or overfilling which could harm the user and or the device 12 Halosil International is not responsible for any injury or damage caused...

Страница 4: ...pen a gallon 3 8 liters bottle of HaloMist Disinfectant and SLOWLY pour the contents into the funnel area AVOID SPLASHING 5 The fluid level indicator lights will illuminate as you pour to show how muc...

Страница 5: ...ensor assume a waiting time of AT LEAST 110 MINUTES AFTER FOGGING 110 minutes is the minimum wait time Length of wait time increases with room size See time chart below Minimum Wait Times Fog Time Rec...

Страница 6: ...om is not blocked c Position the dehumidifier if one is being used outside the smaller interior room centered in the doorway facing towards the fogger d Use Chemical Indicator Strips to ensure that th...

Страница 7: ...d can be connected to the HaloFogger data logger via Bluetooth The HaloLog App does not need to be connected to a fogger before a treatment run is started The on board data logger captures all of the...

Страница 8: ...will be displayed 4 To initiate a connection tap on any HaloFogger serial number displayed on the Halo List Screen When the connection is made the Run Events Screen will appear The top row of the scre...

Страница 9: ...er and total Run Time for that HaloFogger highlighted in GREEN The HaloLog data stored on the selected HaloFogger will download automatically after the Bluetooth handshake has been established which i...

Страница 10: ...vents will identified BILLABLE as a default Events that are not treatments a test run for example can be identified as NON BILLABLE These events will not appear in the downloadable CSV file but will b...

Страница 11: ...address A keyboard will appear to enter address ACTION Tap to email data The iDevice s email screen will open to confirm message ACTION Tap SEND in upper right corner of screen to send message Back u...

Страница 12: ...me to a HaloFogger 2 In the event that HaloLogs of the same HaloFogger have been downloaded to multiple devices If one user has made changes to the HaloLog for example added room building or operator...

Страница 13: ...remove Step 3 Separate Parts Remove Main Bolt from Nozzle Insert Soak both parts in soap and water Rinse parts thoroughly with water Step 4 Reassemble Nozzle Carefully insert the Main Bolt back into...

Страница 14: ...Reservoir Door is not in place Call Customer Service No power The power cord is not plugged in or the power source is not on Check to make sure that the power cord is plugged into both the fogger and...

Страница 15: ...Nozzle Assembly 0750 000 includes retainer cap nozzle insert main bolt and O ring Re Order Number Part Part Number 110V HaloFogger US and Canada 0753 110 220V HaloFogger 0753 220 HaloMist All Purpose...

Страница 16: ...of room being treated CAUTION PRECAUCION DO NOT ENTER ENTERINO Room Fogging in Progress Treatment Date Start Time Estimated Room Re entry Time Refer to HaloMist Use Instructions to determine room re...

Страница 17: ...ion phone and email address purchase date from which HaloFogger was purchased and the unit s serial number Proof of purchase may be required If Halosil International determines that the fogger qualifi...

Страница 18: ...uce infection rates at your facility by visiting www halosil com There you can find out more about the Halo Disinfection System download literature and case studies and find a distributor near you 91...