Halma Perma Pure Zero-Air Скачать руководство пользователя страница 5

 

 
 

 

5

 

COPYRIGHT 2008, ALL RIGHTS RESERVED, Original Instructions 

 PERMA PURE LLC 

 

Zero-Air™ Generator User’s Manual 

Doc. ZA-MAN-001 REV 01 

Page 5 of 10 

 

Specifications and features 

 

Pump: 

 

 

 Dual head diaphragm, up to 30 psig, 18 slpm 

 

Inlet filter: 

 

 Sintered SST, 10 micron porosity 

 

Dryer: 

 

 

 PD-100T-12MZA Nafion

®

 gas dryer 

 

Sterilizer:  

 

 Ultraviolet light chamber, 254 nm peak wavelength 

 

NOx, SOx, HC scrubber: Activated carbon, indicating catalyst blend, 0.7 lbs. 

 

CO catalyst: 

 

 Palladium impregnated alumina 

 

Final filter:  

 

 0.1 micron glass fiber particulate filter  

 

 

Operating environment: -20°C to 40°C ambient temp.; 0-95% RH; Altitude up to 2000 m 

 

Electrical requirements:  115 VAC ±10% 

 

Dimensions:  

 

 8.5” W x 10” H x 16.5” L (216 x 254 x 419 mm) 

 

Weight:   

 

 27 pounds (12.3 kg) 

 
Important safety warnings 

Please be sure to review the following basic safety procedures. These procedures 

represent the MINIMUM requirements to operate the equipment safely. It is the ultimate 

responsibility of the operator to ensure proper safety practices are utilized at the point 

of operation. This equipment is to be installed & operated by trained personnel, with 

sufficient command of the English language to clearly understand the instructions & 

safety warnings. 

 

 

This equipment is NOT designed to operate in a wet environment.

 

 

NEVER operate the equipment with any part of the enclosure unsecured. All operated 

covers must be in place and secured prior to operation. Electrical current may be present 

behind covers, even if tools are not necessary to access these components. 

 

NEVER attempt service on this equipment without first disconnecting all energy sources. 

Repair of this equipment should only be done by properly trained personnel that are familiar 

with the potential risks involved with servicing of the equipment. 

 

NEVER replace fuses with types other then the sample specification of type and current. Do 

not bypass this or any other safety device. 

 

NEVER operate this equipment if it is visibly damaged or the possibility exists that it may have 

been damaged. 

 

The use of components that have not been purchased through an authorized Perma Pure 

dealer or directly from Perma Pure may compromise the safety of the operator. Additionally, 

use of non-authorized components may change the operating characteristics of this equipment. 

Any changes to the equipment, that modify its operation in any way, are dangerous, and are 

strictly prohibited. 

 

Read the entire operating manual before attempting to set up or operate the equipment. 

 

Please heed all warning labels that are on the equipment. They are there to remind you of 

possible hazardous conditions. 

 

Verify the integrity of any mechanical and/or electrical connections that are made to the unit. 

 

Verify that the unit is connected to the proper rated power for the system. 

 

Verify that the unit is plumbed properly to operate effectively. 

Содержание Perma Pure Zero-Air

Страница 1: ...PERMA PURE LLC 732 244 0010 800 337 3762 fax 732 244 8140 www permapure com info permapure com PERMA PURE LLC 1001 NEW HAMPSHIRE AVE LAKEWOOD NJ 08701 Zero Air Generator User s Manual...

Страница 2: ...Introduction 3 General description 4 Specifications and features 5 Important safety warnings 5 Start up procedure 6 Maintenance 7 Activated carbon alumina replacement 7 Final filter replacement 8 Fus...

Страница 3: ...outside of your equipment for any damage If there is any damage contact the carrier immediately Generally a carrier must be notified within 24 hours of the delivery to make a hidden damage claim Items...

Страница 4: ...urrent purge air Ultraviolet Lamp The UV lamp produces approximately 1 3 ppm of ozone which breaks down microorganisms into carbon compounds that are removed further downstream It also oxidizes NO to...

Страница 5: ...closure unsecured All operated covers must be in place and secured prior to operation Electrical current may be present behind covers even if tools are not necessary to access these components NEVER a...

Страница 6: ...r clockwise to open fully 5 Connect calibration gas line to tube compression fitting located on top panel of generator 6 Turn power switch to ON position 7 Adjust flow control valve to desired air out...

Страница 7: ...o brown When of the canister has changed to brown replace mixture The mixture can not be regenerated and must be discarded Replacing the mixture 1 Place towel or cloth across bottom panel to prevent s...

Страница 8: ...se replacement If unit fails to run when plugged in and turned on first check the fuse 1 Disconnect power cord from power input module on bottom panel of unit 2 Using slotted tip screwdriver pry up at...

Страница 9: ...r that the product shall in fact be suitable or desirable for any particular use or application 2 Buyer takes sole responsibility for the use and applications to which the product is put and Buyer sha...

Страница 10: ...PERMA PURE LLC Zero Air Generator User s Manual Doc ZA MAN 001 REV 01 Page 10 of 10 Appendix B Wiring diagrams Figure 4 Wiring diagram 115 VAC...

Отзывы: