1: Setup
Connecting a Spectrometer to the HL-2000-CAL
►
Procedure
Use the following procedure to connect your HL-2000-CAL unit to a spectrometer:
1. Loosen the set screw on the HL-2000-CAL's SMA connector
2.
If using a bare fiber with the lamp
:
Remove the inner barrel from the SMA connector (if using a bare fiber with the lamp). Screw this
connector barrel onto the end of your fiber. The connection should be tight. Insert the barrel/fiber
all the way into the HL-2000-CAL's SMA connector.
If using a fiber and cosine corrector with the lamp:
Remove the inner barrel from the SMA connector (if using a fiber and cosine corrector with the
lamp). Screw the CC-3 cosine corrector onto the end of the fiber. The connection should be tight.
Insert the CC-3/fibcr all the way into the HL-2000-CAL's SMA connector, taking the place of the
SMA's inner barrel.
3. Tighten the setscrew on the SMA connector of the HL-2000-CAL with an Allen wrench. Screw
the other end of the fiber into the SMA connector of the spectrometer.
4. Plug the switching AC adapter (the black rectangle-shaped box) into the back of the HL-2000-
CAL. This adapter stabilizes the power coming into the lamp to insure constant spectral intensity.
5. Plug one end the power cord into a standard outlet, then plug the other end of the power cord into
the back of the switching AC adapter.
6. Turn the lamp on. Allow the lamp to warm up for at least 15 minutes before using it.
Configuring the OOIIrrad Software
►
Procedure
Use the following procedure to configure the OOIIrrad software that comes with your HL-2000-CAL:
1. Insert the disk that came with the HL-2000-CAL lamp into the computer.This disk contains two
ASCII files containing the same information as the Lamp Calibration Reports that came with the
HL-2000-CAL:
•
One file has the calibration numbers for calibrating the spectral response of your system with
the lamp and a
bare fiber
.
The file name contains the lamp serial number, followed by
FIB.LMP.
000-10000-060-02-0505
3
Содержание Ocean Optics HL-2000-CAL
Страница 4: ...Important Safety Notices B 000 10000 060 02 0505...
Страница 6: ...Table of Contents ii 000 10000 060 02 0505...
Страница 14: ...2 HL 2000 CAL Specifications 6 000 10000 060 02 0505...
Страница 16: ...3 Calibration 8 000 10000 060 02 0505...
Страница 18: ...Index 10 000 10000 060 02 0505...