background image

 

 

 

 

 

 

 

 
 
 
 
 

DH

-

mini UV

-

Vis

-

NIR Deuterium

-

Halogen 

 

Light Source with Shutter

 

 

Installation and Operation Manual

 

 
 

 

 
 

 

 
 
 
For Products: DH-mini UV-Vis-NIR 
Document: 000-10000-200-02-201510 

 

   

 

 

        

 

 

 

 

Содержание Ocean Optics DH-mini UV-Vis-NIR

Страница 1: ...DH mini UV Vis NIR Deuterium Halogen Light Source with Shutter Installation and Operation Manual For Products DH mini UV Vis NIR Document 000 10000 200 02 201510...

Страница 2: ...operty of their respective owners Limit of Liability Every effort has been made to make this manual as complete and as accurate as possible but no warranty or fitness is implied The information provid...

Страница 3: ...ction against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions...

Страница 4: ...Important Safety Notices B 000 10000 200 02 201510...

Страница 5: ...Setup 1 Overview 1 Unpacking the DH mini 1 Contents 1 Attaching to Ocean Optics Spectrometers 2 Setting Up the DH mini Light Source 3 Chapter 2 Operation 5 Overview 5 Operating Modes 5 Manual Mode 6...

Страница 6: ...Table of Contents ii 000 10000 200 02 201510...

Страница 7: ...ell as changing the intensity of the Halogen bulb Appendix A Specifications Contains operating environment specifications as well as other physical details of the product Appendix B Bulb Replacement P...

Страница 8: ...comprehensive warranty ensures you of the highest level of craftsmanship and reliability for years to come No other manufacturer offers such a solid guarantee of quality and reliability The Ocean Opti...

Страница 9: ...on the unit and in this manual Unpacking the DH mini Procedure 1 Unpack your lamp assembly carefully Dropping this instrument can cause permanent damage 2 Inspect the outside of the instrument and ma...

Страница 10: ...nd pinouts for attaching the DT mini light source to various Ocean Optics spectrometers DB15 Pin DH Mini HR QE Maya NIRQuest Flame STS USB2000 Cables Adapters Needed N A HR4 CBL DB15 FLAME CBL DD4P DB...

Страница 11: ...wing procedure to set up your light source Procedure Perform the steps below to set up your DH mini Light Source 1 Plug the power supply into a wall outlet Power Supply 2 Plug the other end of the pow...

Страница 12: ...1 Setup 4 000 10000 200 02 201510 3 Connect the SMA connector of your fiber optic cable to the SMA plug on the front of the light source Note For UV measurements a solar resistant fiber is recommended...

Страница 13: ...LED will glow and the device will be in TTL control mode which is highlighted by the yellow TTL Manual LED Operating Modes To toggle between Manual Mode and TTL Mode you have to push the TTL Manual bu...

Страница 14: ...urements Press button again to turn the Halogen bulb off The LED extinguishes Shutter Press button to open shutter The LED lights Press button again to close the shutter The LED extinguishes TTL Manua...

Страница 15: ...ntrol 5 V bulb on 2 Halogen control 5 V bulb on 3 Shutter control 5V shutter open 4 GND There is a TTL adapter cable available to directly attach the DH mini on an Ocean Optics spectrometer using a HR...

Страница 16: ...trometer you can use the Strobe Lamp Enable check box to control the shutter See Attaching to Ocean Optics Spectrometers for more pinout information Change Intensity of Halogen Bulb To change the inte...

Страница 17: ...ter 30 minutes stabilization Warm up time 6 minutes at 20 C Power consumption 12 W Optical fiber connection SMA 905 Shutter input TTL input up to 2 5 Hz maximum Pin position at SUB D 15 pin adapter al...

Страница 18: ...spectrometers have measured irradiance spectra The balance from the Deuterium spectrum 200 400nm to the Halogen spectrum 400 1600 nm can be changed by varying the intensity of the Halogen bulb see Ch...

Страница 19: ...separately To order replacement bulbs for the DH mini see Spectral Output Replacing the Bulb Procedure To replace the bulbs 1 Remove all eight countersunk screws on each side of the light source hous...

Страница 20: ...B Bulb Replacement 12 000 10000 200 02 201510 4 Pull the bulb from the light source 5 Insert a new bulb Be careful to correctly align it with the connection holes on the PCB 3...

Страница 21: ...000 200 02 201510 13 6 Replug the cable connection for the Halogen bulb on the rear side of the replacement bulb The polarity of the connector doesn t matter 7 Replace the housing and secure the housi...

Страница 22: ...B Bulb Replacement 14 000 10000 200 02 201510...

Страница 23: ...ntensity 8 M Manual mode 6 O OceanView 8 operating modes 5 Manual 6 TTL 6 operation 5 P package contents 2 parts list 10 product related documentation iii S set up 3 setup 1 specifications 9 spectral...

Страница 24: ...Index 16 000 10000 200 02 201510...