SERIAL-BT-ADP Getting Started Instructions
9600 Baud Rate Switch Settings
This is the default baud rate for most Ocean Optics spectrometers: Maya2000 Pro, NIRQuest, QE65000,
STS, Torus, and USB4000.
Unit 1 (master) switch settings:
9.6K/115K baud
Switch 4 = on
100-40000-032-01-201203b
3
Auto Master
Switch 3 = on
Auto Discover
Switch 2 = on
Factory Defaults
Switch 1 = off
Unit 2 (slave) switch settings:
9.6K/115K baud
Switch 4 = on
Auto Master
Switch 3 = off
Auto Discover
Switch 2 = on
Factory Defaults
Switch 1 = off
These setting will set the Bluetooth pair to 9600 baud, no parity, 8 bits, 1 stop bit, 1 start bit. These
settings will also cause unit 1 to act as a master and search for the nearest slave (unit 2). It is important
that no other slave devices are closer than the unit 2.
Note
Once the SERIAL-BT-ADP devices have been successfully paired, set Switch 2
(Automatic Discovery) to off to prevent them from inadvertently pairing with any other
Bluetooth devices within range.
115200 Baud Rate Switch Settings
This is the default baud rate for the following spectrometers: and HR4000.
Unit 1 (master) switch settings:
9.6K/115K baud
Switch 4 = off
Auto Master
Switch 3 = on
Auto Discover
Switch 2 = on
Factory Defaults
Switch 1 = off
Unit 2 (slave) switch settings:
9.6K/115K baud
Switch 4 = off
Auto Master
Switch 3 = off
Auto Discover
Switch 2 = on
Factory Defaults
Switch 1 = off
These setting will set the Bluetooth pair to 115200 baud, no parity, 8 bits, 1 stop bit, 1 start bit. These
setting will also cause unit 1 to act as a master and search for the nearest slave (unit 2). It is important
that no other slave devices are closer than the unit 2.
4
On
Off
3
2
1
4
On Off
3
2
1
4
On
Off
3
2
1
4
On
Off
3
2
1