Ensure the flow path is the
same direction as the flow
arrow shown on your unit.
DOC-MANUAL-MB-QUICK-US
Rev 1 • 2021-06-02
Get Started:
This box includes:
• Alicat device
• NIST traceable calibration certificate
• Power and communication cable
• Optional accessories
(filters, fittings)
Power
The battery is shipped
partly charged. Once
fully charged, it lasts
up to 18 hours. Use the
supplied power cable
with an outlet adapter
for fastest charging.
Gas and Units
Select your gas:
MENU
→
SETUP
→
ACTIVE GAS
. Ensure the
proper units are selected:
Press
MENU
→
SETUP
→
SENSOR
→
ENGINEERING UNITS
.
Tare
Before installing
the device into your
process, press the
TARE FLOW
button
on the main screen.
This will zero out
(tare) your unit.
Model
Series
Maximum
line pressure
Maximum
differential pressure
MB, MBS
145 PSIG
75 PSID
MWB
45 PSIG
10 PSID
MBQ
305 PSIG
75 PSID
You’re Ready to Connect:
Download the full manual:
Alicat.com/manuals
Annual Calibration • Troubleshooting • Questions:
+1 520-290-6060 • [email protected] • Live web chat at Alicat.com
5
Verify the requirements for
pressure above, and connect
the device to your process.