CODA-Series Coriolis mass flow meter operating manual 

11

Multidrop Information

CODA devices equipped with an RS-485 interface can operate on 

networks with other devices. The CODA RS-485 transceiver is a 1/2 unit-

load transceiver. When deploying CODA devices in an RS-485 network 

with multiple types of devices be sure to confirm that the total load of 

all devices does not exceed 32 unit loads on an unrepeated network 

segment. Consult the EIA-485 standard for more information. 
CODA KM-Series devices equipped with an RS-232 interface do not have 

a multidrop capable RS-232 transceiver.
If a CODA device is plugged into an RS-232 network with other devices 

which have been adapted to support multidrop communication over 

RS-232 it will load the bus excessively and prevent other devices from 

communicating on that network. This includes other Alicat M-Series, 

L-Series, and P-Series devices.

Modbus RTU Serial Protocol

Alicat uses the Modbus standard in which registers are offset by 1 from 

addresses, meaning register 1 is equivalent to address 0, but some 

systems expect data to be mapped as 0-indexed addresses. Different 

Modbus control systems may refer to registers, offsets, or addresses 

in their documentation without clarifying their meaning. If your control 

system uses a 0-indexed numbering scheme then decrement all register 

references in this manual by 1. 
If you are unsure of which addressing scheme your control system uses, 

perform a test read of register 1200. If the CODA instrument responds with 

Error code 2: “Illegal Data Address”, then your system is using the standard 

1-indexed numbering system and the values in the manual can be used 

as-is. If the device returns a value of 0 instead of an error, decrement all 

registers listed in this section by 1 to arrive at the correct offset.

Содержание Alicat Scientific CORIOLIS CODA KM Series

Страница 1: ...OPERATING MANUAL CODA SERIES CORIOLIS MASS FLOW METER Model KM...

Страница 2: ...Hong Cao Rd Shanghai 200233 PRC 86 21 60407398 ext 801 Europe europe alicat com Geograaf 24 6921 EW Duiven The Netherlands 31 0 26 203 1651 This device comes with a NIST traceable calibration certific...

Страница 3: ...ion 9 Analog Communication 9 Analog Output Electrical Characteristics 9 Analog I O Data Ranges 10 Serial Communication 10 Establishing Communication 10 Multidrop Information 11 Modbus RTU Serial Proto...

Страница 4: ...ounting holes on the bottom for convenient attachment to flat panels No straight runs of pipe are required upstream or downstream These devices are position insensitive and can be mounted in any orien...

Страница 5: ...ODA Series Coriolis mass flow meter operating manual 5 1 4 VCR Inlet Male DB 15 Connector Tare Button LED Power Indicator USB Type C 1 4 VCR Outlet 1 024 0 207 1 731 0 914 Mounting Holes 2 M5 0 8 10mm...

Страница 6: ...Operating the device above the proof pressure can permanently damage the device Operating above the burst pressure for even short periods will result in the device rupturing High Pressure variants of...

Страница 7: ...ed 0 5 Vdc output of mass flow rate Optional 0 10 Vdc or 4 20 mA 2 Not connected Power in 3 0 5 Vdc output of mass flow rate Optional 0 10 Vdc or 4 20 mA Serial RS 232 RxD Optional RS 485 A 4 0 5 Vdc...

Страница 8: ...onds to initiate the 10 second taring process page 7 Tare via Serial CODA KM Series meter can be tared by pushing the Zero Device button in Alicat s CODA View program page 15 Once this button is press...

Страница 9: ...ave a single analog output for mass flow page 7 Both of these outputs are linear across the entire range provided the load impedance is within the nominal values specified in the table below Analog Ou...

Страница 10: ...gital serial interface in addition to the analog interface Modbus RTU is the default communications protocol for KM Series meters You can read and log sensor data and adjust device settings via these...

Страница 11: ...municating on that network This includes other Alicat M Series L Series and P Series devices Modbus RTU Serial Protocol Alicat uses the Modbus standard in which registers are offset by 1 from addresse...

Страница 12: ...d and control requests to a CODA instrument are issued with Modbus function code FC16 write multiple registers For PID tuning assistance please contact Alicat Modbus Reading and Status Registers Note...

Страница 13: ...air is issued as a set of two 16 bit unsigned integers Commands are started by a write to register 1000 If you send a command to register 1000 without sending an argument to 1001 the CODA instrument w...

Страница 14: ...zer 5 Reset totalizer value 0 Pause totalizer 1 Resume totalizer 15 Pause totalizer Requires a power cycle to take effect 1 247 New ID 32767 Change Modbus ID Requires a power cycle to take effect 9600...

Страница 15: ...ng the CODA View program select the COM port your device is connected to in the top left corner and click Connect If connecting via the USB C connection on the CODA instrument please ensure that the c...

Страница 16: ...an effort to remove all of the liquid from the device as deposits of calcium or other soluble minerals can affect the accuracy of the device Recalibration CODA instruments are calibrated to NIST trac...

Страница 17: ...s covered by this warranty Alicat Scientific does not recommend warrant or assume responsibility for the use of the Products in life support applications or systems Alicat s warranties as herein above...

Страница 18: ...Ltd Plot NO A 147 Road No 24 Next to Spraytech Circle opp Metropolitan Company Wagle Industrial Estate Thane West Mah r shtra 400 604 1 888 290 6060 China SE Asia info cn alicat com alicat com cn 2nd...

Отзывы: