1
Get Started:
This box includes:
• CODA Coriolis mass flow controller
• NIST-traceable calibration certificate
• Optional accessories
(cables, power adapters, filters)
Install
Install the device into
your process. Ensure the
flow path is the same
direction as the flow
arrow shown on your
unit. The device may be
oriented in any direction.
Signal Output
See reverse side for pinout details
Analog:
Signal outputs are
available via 0–5 Vdc (0–10
Vdc or 4–20mA options)
Digital:
USB-C, RS-232,
RS-485 and MODBUS RTU
over RS-232 or RS-485.
DOC-MANUAL-KC-QUICK-US
Rev 0 • 2020-11-05
Download the full manual:
Troubleshooting • Questions:
+1 520-290-6060 • [email protected] • Live web chat at alicat.com
Power
Supply power via the
DB-15 connector. The
controller requires
12–30 Vdc. A green
LED light indicates
power input.
Analog:
Apply a
ground to Pin 12 for
at least 1 second.
(See reverse side for
full pinout details.)
Digital:
For serial
tare options including
MODBUS and
LabVIEW, consult
the full manual.
Button:
Press and
hold the tare button
on top of the device
for 10 seconds.
2
3
Zero (Tare)
After completing step 5, tare the device with one of three methods:
4
Although the device is
immediately functional,
allow for 15 minutes of
warm-up for best accuracy.
5
Flow the fluid to be measured
through the device, then estab-
lish a no-flow condition while fluid
is present throughout the unit.
6
1
12
Ground
to tare
DB-15 Male