Hallowell EMC 2002PRO Скачать руководство пользователя страница 9

2/26/2018 

DOCA3667B Pro Series Operating Manual.doc 

9 of 22 

SET-UP PROCEDURE (Continued) 

 
8.

 

Insert the Airway Pressure Sampling Tee into the breathing system. 

Disconnect the 

patient breathing hose from the INHALE VALVE of the anesthesia machine. Connect the 
Airway Pressure Sampling Tee (PN 000A2420B) to the INHALE VALVE and reconnect 
the patient breathing hose to the Airway Pressure Sampling Tee. Route the sampling tube 
as desired to the ventilator Pressure Transducer port on the rear panel. Trim sampling tube 
to length leaving enough slack for movement. Install Luer Lock fitting such that the hose 
barb is fully set into the tube. Attach fitting to ventilator Pressure Transducer port. 

 
9. 

Connect the ventilator to the scavenger.

  Use a 19-mm corrugated tube (not provided) to 

connect the EXHAUST port of the bellows assembly to a properly functioning scavenger 
system. 

 

Warning:

 

Applying any negative or positive pressure to the EXHAUST port of the bellows base assembly 
will result in a more positive pressure in the patient breathing system and improper operation of 
the ventilator. 

 
10. 

Connect the supply gas.

  Connect the supply gas hose to the 50 psi SUPPLY GAS port of 

the control unit, and to the supply gas source.  The unit is provided with a DISS 1240 male 
oxygen bulkhead fitting.  Gas consumption of the ventilator is very economical, therefore, 
the use of oxygen as a drive gas is recommended.  Oxygen use reduces the risk of unit 
malfunction due to contamination of the pneumatics.  Compressed air may be used, but it 
must be CLEAN and DRY. 

 

The 50-psi source for the ventilator may be provided from a separate tank, wall or ceiling 

drop, or from a PTO (Power-Take Off) on the anesthesia machine.  Should none of the 
above be available, a common practice, for machines with DISS 1240 male 

O

2

 inlets, is to 

remove the 

O

2

 supply line to the anesthesia machine from the anesthesia machine, connect a 

demand wye (HEMC PN 150A1691 (GRN) or 150A1692 (WHITE)) to the inlet of the 
anesthesia machine and then reconnect the 

O

2

 supply line to the anesthesia machine via one 

of the two remaining connections to the wye.  This leaves one leg of the wye available to 
source the ventilator with 

O

2

, connect HEMC PN 000A0489 (GRN) or 000A0490 (WHITE) 

between the demand wye and the ventilator 50-psi inlet. 

 
11. 

Connect the electrical power.

  PRIOR to connecting the electrical power, confirm that the 

power inlet module voltage selector switch is set appropriately for your location. 

 

Export Customers Beware 

Warning: 

Turning the ventilator on with the Power Inlet Module Voltage Selector switch set for 110 volts 
while the unit is plugged into a 208 - 240-volt source WILL damage the unit and void the 
warranty.  

 

Special order 100Vac Japanese versions require no voltage selector setting and will work 
ONLY on 100Vac. 
 
Plug the ventilator into a properly grounded power source. 
 

Содержание 2002PRO

Страница 1: ...Hallowell EMC Model 2002PRO Model 2002IEPRO Veterinary Anesthesia Ventilators OPERATING SET UP MANUAL...

Страница 2: ...CAUTIONS 3 WARRANTY 4 5 INTRODUCTION 6 RECEIVING PROCEDURES 7 SET UP PROCEDURE SET UP DIAGRAM 8 9 10 CONTROLS 11 ALARMS 12 VERIFICATION OF PROPER FUNCTION 12 13 GETTING STARTED 14 15 OPERATING INSTRU...

Страница 3: ...ons unauthorized repair or modification of the equipment or accessories or from the use of components or accessories that have either been damaged or not authorized for use with this equipment by Hall...

Страница 4: ...ited warranty WHAT THIS WARRANTY DOES NOT COVER 1 Defects or damages resulting from use of the product in other than its normal and customary manner 2 Defects or damages from abnormal use abnormal con...

Страница 5: ...low the exclusive of limitation of incidental or consequential damages or limitations on how long an implied warranty lasts these limitations or exclusions may not apply to you This warranty gives you...

Страница 6: ...impact resistant material that is unaffected by water or the presence of all state of the art anesthetics In addition the high temperature properties of ULTEM make it fully compatible with standard s...

Страница 7: ...ng Model 2002PRO 2002IEPRO Controller Bellows Base Assembly PN 000A0510 Tube Driving Gas 7 black rubber PN 000A0495 Airway Pressure Sampling Tee PN 000A2420B 36 x 22 mm Breathing System Tube PN 201A16...

Страница 8: ...unting ring Carefully hold the outer edge of the bellows disk top of bellows lift and lower it quickly several times to puff out and remove any folds in the convolutions Place the bellows housing over...

Страница 9: ...r is very economical therefore the use of oxygen as a drive gas is recommended Oxygen use reduces the risk of unit malfunction due to contamination of the pneumatics Compressed air may be used but it...

Страница 10: ...2 26 2018 DOCA3667B Pro Series Operating Manual doc 10 of 22...

Страница 11: ...ntrol is described in detail below I O Power Switch a toggle switch used to turn the ventilator on The green LED indicates the power is on when illuminated INSPiratory HOLD Control A momentary pushbut...

Страница 12: ...sor for this alarm is located downstream of the internal 40 micron filter and may also indicate a clogged filter condition During an Alarm condition the yellow light on the front panel is illuminated...

Страница 13: ...the breathing system is not leaking The pressure should remain constant Release the INSPiratory HOLD button Observe that the ventilator is continuing to cycle and that over time ten cycles and the bel...

Страница 14: ...the maximum working pressure limit MWPL control to about 20 cm H2O The airway pressure will not exceed this setting regardless of what you do with the other controls Connect the ventilator to the bre...

Страница 15: ...ge being implemented you have changed the MV as desired now If needed trim the delivered TV size with the RATE control to obtain a new TV that results in a more appropriate PIP and no alarms Now do it...

Страница 16: ...ppropriate RATE has been selected Turn the VOLUME controls counterclockwise to increase the tidal volume delivered or clockwise to decrease the tidal volume delivered Read the approximate tidal volume...

Страница 17: ...a tight o ring fit Tilt the top of the bellows housing toward you and lift it off DO NOT attempt to steam sterilize the bellows housing Since it does not come in contact with the breathing gas it nee...

Страница 18: ...the base and loosen the four black thumbscrews located at the corners First clean the base using a clean soft slightly dampened cloth Then wrap the whole base and steam sterilize it using the same sta...

Страница 19: ...produce a closed loop 4 While observing the breathing system pressure slowly increase the fresh gas flow to the system until the pressure builds to about 50 cm H2O 5 Turn the flow off 6 The pressure s...

Страница 20: ...llapsed bellows Unkink the drive gas tube Unkink breathing system hose The ventilator operation sounds normal but the TV delivered is incorrect and or inconsistent Missing or damaged pop off valve o r...

Страница 21: ...name address phone number and PO number of the responsible party 5 If the item is to be returned such as with a repair the name of the party to whom we should ship the item the shipping address and r...

Страница 22: ...mum Working Pressure Limit Linearly Adjustable 10 60 cm H2O Power On Front Panel mounted green LED Alarm Visual Front Panel mounted yellow LED Alarm Audio Internal audio transducer Alarm Audio Int Low...

Отзывы: