Shelf
Door handle
Control
pannel
Battery switch
Inner light
Casters
Function and Instruction For 2-8
Pharmaceutical
Refrigerator
5 Adjust alarm temperature range: See fig #1
6 Defects display
7Inner light
Two methods set high temperature alarm “HA” and low temperature alarm “LA”
Press “Alarm” button for 10 seconds. LED display code “HA”,press “SET” button again(quickly),code
“HA” flash, adjust the set- temperature by both pressing “+” and
” buttons and increment 1°C,
adjustment range shouldn't be 6°C higher than pre-set cabinet temperature,press “Alarm” button to
return the former menu(each press of “Alarm” button will return one step back to former menu until
display cabinet temperature) and LED display “HA” Press both “+” and
” to choose another
parameter “LA”, “LA” setting method like “HA”, and its adjustment range shouldn't be 6° C lower than
pre-set cabinet temperature.
The control panel only can display one sensor error, when the sensor error happens.LED display
“E1”,and alarm at the same time. The refrigerator will keep working according to the defect safty
model,and you should put the stored goods to other refrigerator as soon as possible,and cantact the
after-sales department quickly.
In order to see the stored goods clearly, a LED light has been installed in the cabinet. You can control
light “on” or “off”by “light” button to save energe,please turn off the light after observing the stored
goods in the refrigerator.
The light is on when the door is open; The light is off when the door is closed.