Function and Instruction For 2-8

Pharmaceutical

Refrigerator

5 Adjust alarm temperature range: See fig #1

6 Defects display

7Inner light

Two methods set high temperature alarm “HA” and low temperature alarm “LA”

Press “Alarm” button for 10 seconds. LED display code “HA”,press “SET” button again(quickly),code

“HA” flash, adjust the set- temperature by both pressing “+” and

” buttons and increment 1°C,

adjustment range shouldn't be 6°C higher than pre-set cabinet temperature,press “Alarm” button to

return the former menu(each press of “Alarm” button will return one step back to former menu until

display cabinet temperature) and LED display “HA” Press both “+” and

” to choose another

parameter “LA”, “LA” setting method like “HA”, and its adjustment range shouldn't be 6° C lower than

pre-set cabinet temperature.

The control panel only can display one sensor error, when the sensor error happens.LED display

“E1”,and alarm at the same time. The refrigerator will keep working according to the defect safty

model,and you should put the stored goods to other refrigerator as soon as possible,and cantact the

after-sales department quickly.

In order to see the stored goods clearly, a LED light has been installed in the cabinet. You can control

light “on” or “off”by “light” button to save energe,please turn off the light after observing the stored

goods in the refrigerator.

The light is on when the door is open; The light is off when the door is closed.

Содержание HYC-68

Страница 1: ...68 68A Operation Manual Read the Operation Manual carefully before using your appliance Keep the Operation Manual in a safe place Appearance color and layout of the door may vary Pharmaceutical Refrigerator ...

Страница 2: ...10 11 12 10 HYC 68 HYC 68A 68L 68L Insulation door Glass door 2 2 R600a 32kg 32kg HYC 68 HYC 68A 84W 84W ...

Страница 3: ......

Страница 4: ...e appliance is first connected to the power source and can be inactivated by pressing the buzzer buttom The warning indicator stays on until the temperature of the test bottle sensor reaches 5 2 3 Ensure that the two test bottles are filled with 10 glycol solution see article 4 section installation on page 6 4 The appliance is preset during manufacture for operation temperature 2 8 ...

Страница 5: ...8 ...

Страница 6: ... until display cabinet temperature and LED display HA Press both and to choose another parameter LA LA setting method like HA and its adjustment range shouldn t be 6 C lower than pre set cabinet temperature The control panel only can display one sensor error when the sensor error happens LED display E1 and alarm at the same time The refrigerator will keep working according to the defect safty mode...

Страница 7: ...ure or lower than the low alarm temperature buzzing and light flash Press Alarm button to end the buzzer but light flash going on until the cabinet temperature renew to the alarm temperature range Press both Alarm and button to check high temperature alarm loosen buttons to return cabinet temperature display Press both Alarm and button check low temperature alarm loosen buttons to return cabinet t...

Страница 8: ...For more information please refer to the cleaning section After placing the refrigerator flat and completing the cleaning do not electrify the refrigerator until after 30minutes of standstill so as to ensure smooth operation of the appliance Remove all package assembly Check off attached accessories and data Place the refrigerator on a suitable position Clean the refrigerator before operation Stan...

Отзывы: