Temperature adjusting:

Press and hold both SENSOR SELECT button and CALIB CANCEL button for more than 3s until

original setting starts flashing to indicate that it enters temperature setting mode. With each press

of CALIB CANCEL button, temperature setting increases 0.1

until reaching upper limit, which

is defaulted as 8

. If press again, it returns to lower limit, which is defaulted as 2

. It repeats the

above cycle. When it reaches your desired temperature, if you don't press the button within 5s,

the temperature setting will be saved and it automatically quits the temperature setting mode (the

temperature setting range programmed at factory is 2 - 8

). The display returns to display inside

sensor temperature or average temperature before setting is made (not flashing).

2~8

.

8

in side

Отзывы: