![Haier HYC-1378 Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/haier/hyc-1378/hyc-1378_operation-manual_2101638005.webp)
26
3
Safety Precautions
Dear users:
Thanks for your choice of Haier pharmaceutical refrigerator; please make sure you have carefully
read and observed the contents with following signs in the manual, for better understanding of this
manual and better use of this product, so as to prevent personal injuries and refrigerator damage.
Ignoring this warning may result in
death or serious injury
Ignoring this warning may result in
death or serious injury,and/or damage
to the refrigerator and property
Actions or operations
which are prohibited
Actions or operations
which must be followed
If the power line needs to be extended, the extended line shall be no smaller than 2mm² in
sectional area and no longer than 3m in length. Otherwise, fire or electric shock may be
caused.
The power line for this refrigerator is provided with a standard three wire (grounding) plug,
which complies with 16A. Do not remove or dismantle the grounding pin of power line in
any case. Make sure the power plug is securely and reliably plugged into the socket to
prevent fire.
Apply socket with ground wire to prevent electric shock.If the socket fails to be grounded,
ground wires must be installed by professional technical staffs.
The refrigerator shall not conduct outdoor service. Electrical leakage or shock may be
caused if wet by rainwater.
Do not place the refrigerator in humid places or places where the refrigerator may suffer
splashing water, to prevent electric leakage or shock due to deterioration of insulation.
Do not pour water on the refrigerator to prevent electric shock or short circuit.
In case of flammable gas leakage such as coal gas, shut off the valve of leaked gas, open
windows for ventilation and exhaust; do not plug in or pull out the power plug of refrigerator
to prevent explosion and fire.
The refrigerator shall be installed by professional technical staffs or after-sale maintenance
staffs to prevent electric shock or fire.
Place the refrigerator on solid and flat ground in a stable manner. The refrigerator will be
tipped over or personal injuries will be caused if the refrigerator is placed on improper
ground or place.
Apply the dedicated power supply marked on the nameplate of refrigerator to prevent fire
or electric shock.
If the voltage being used is 10% higher or lower than rated voltage , automatic voltage
regulator above 4000W and appropriate for motor load shall be installed.
1440×910×1980
1320×700×1500
1378L
Pharmaceutical Refrigerator
Outside Dimensions
(W×D×H)
(mm)
Inside Dimensions
(W×D×H)
(mm)
Effective Volume
Door
Foaming Material
Compressor
Shelf
Refrigerating Method
Outer Plate /Inner Plate
Condenser/Evaporimeter
Refrigerant
Lamp
Net Weight
Temperature of Monitoring Bottle
Rated Power Supply
Rated Power/Current
Alarm Equipment
High & low temperature alarm, sensor fault alarm,
power failure alarm, Low battery alram, door opening alarm
Name
Model
HYC-1378
R134a 490g
310kg
2
℃
~8
℃
Duration of Outage Alarm
48 h (when the battery is fully charged)
Type of Electric-shock Safeguard
I
220V-240V~/50Hz
800W/4.2A
Temperature Controller
Specification
Specification
Glass door with electric heat
Rigid polyurethane foam (fluoride-free)
Fully enclosed compressor
12
Force-air cooling circulation
Cold-rolled steel with power cover
Finned pipes / Finned pipes
Computer control system
LED 5W
Power connection type
Y
Climate Type
4
Foaming Agent
CP/IP
Noise Level
43dB(A)
The loading of each shelf
Max:30kg/m²
Note: Climate Type 4 means the temperature is +30
℃
,the relative humidity is 55%.
Testing hole
2
R134a 490g
220V-240V~/60Hz
800W/4.2A
R134a 490g
115V~/60Hz
950W/11A
48dB(A)
Rechargeable Battery
DC 12V rechargeable battery