Haier HXC-1308 Скачать руководство пользователя страница 24

Change Recording Paper 
The record paper must be replaced by a new piece of paper after a record period, and see the 
steps of changing record paper below: 
Press Key 3#, the scriber will move to the outer side; 
After the scriber has moved away from the record paper totally, release the hold-down bolt and 
change new record paper, and please align the starting time scale of record paper with the 
reference line of panel carefully; 

The adjustment modes of temperature measurement tap position and record period. 
After normal power-on of machine, press and hold Key 3# for at least about 5s, when a “click” is 
heard, the pointer of recorder stops walking, which means the recorder switches to the set mode; 
Press Key 1# to switch over temperature measurement tap position and the color of indicator
 light shown in following figure will change with each press. Different colors represent different tap 
positions, and the corresponding of color and tap position is as follows: 
Green: -40°C~+30°C (applicable to this product and must be selected) 
Yellow: -115°C +50°C 
Red: -200°C

+0°C 

Note: According to aforementioned temperature tap position, before the temperature reaches the 
record scope, the recorder won’t work. 
Press Key 2# to select record period. There are two kinds of record period in total to choose and 
they should be distinguished according to the sound made by machine: 
The record period is 24h: rapid, short buzzing made by machine; 
The record period is 7 days: Continuous short buzzing made by machine: (recommended period 
and saving paper) 
Adjust tap position and period in place and press Key 3# to exit setup. 
Paper calibration 
After power-on, tap position and period adjustment, for the accuracy of test, the setting of tap 
position and period will take effect after paper calibration on equipment. If tap position and period 
are not adjusted after power-on, 5s later, the machine will enter paper calibration mode 
automatically and the paper calibration mode is as follow: 
After pressing 3#, the scriber will move to the mark position of -40t firstly and stay here for 3s. 
check whether pointer position and the scale of paper are matched at this time. If not, the pointer 
can be adjusted finely by pressing 1# or 2#. 
3s later, the scriber will: Move to the position of -20°C from that of -40°C. And stay here for 3s. At 
this time, the pointer also can be adjusted finely according to aforementioned method; 
Then the scriber will move from the position of -20

 to scale position of 0

 to be measured and 

stay here for 3s, and the pointer position can also be adjusted finely; 
3s later, the scriber will slide to appointed scale according to the present perceived temperature, 
then the calibration ends. 

After retightening hold-down bolt and pressing Key 3# again, the recorder will switch to the 
process of paper calibration. Please refer to the explanation of last page to calibrate the 
installation of paper; 
After calibration, the temperature recorder will switch to normal record status. 

Time and ID (serial no. of recorder) setting of recorder 
Insert the storage card of recorder into the recorder, the indicator light of storage card flashes. 
Press Key B, if the time and ID of storage card has been set, the recorder will update the time 
and ID, and eliminate all internal records such as temperature and time record. 
After operation is complete, the indicator light of storage card will be on constantly, otherwise the 
light keep flashing status; 

Data reading of recorder 
Insert the storage card of recorder into the recorder, the indicator light of storage card begins to 
flash; 
Press Key A, if there is enough space in storage card, the data will be transferred into storage 
card, and the indicator light will be on constantly after successful transfer. If the data fails to be 
transferred, the indicator light of storage card will maintain flashing status. 
Remove the storage card and connect it with computer to view data through equipped USB kit. 

     Calibration adjustment of recorder 
Connect the temperature recorder to power supply and put temperature sensing probe in a stable 
temperature environment (e.g. 0

 of ice water mixture), after the temperature recorder enters 

temperature status, press and hold 1# or 2# (for about 5s), the recorder will enter temperature 
calibration status after a “click” made by buzzer. Apply 1# and 2# to conduct lateral adjustment to 
make the pointer align with the temperature scale on record paper, stop adjustment, press Key 3# to 
confirm or for 5s, after a “click” made by buzzer, the temperature recorder will exit temperature 
calibration status automatically and return to temperature record status. 

The recorder saves data in storage card every 1h and can keep 2 weeks of 
data at most. When the usable space is full, the new data will cover the earliest 
hour of record. 
After the storage card has read out data, the recorder will still save the collected 
data. Only when ID or real time is reset, the data can be cleared. 

The recorder has been calibrated before leaving factory. Please do not adjust 
the recorder unless special circumstances. 
When the refrigerator is not energized and discontinued, please be sure the 
key of recorder is in the “0” end, avoid over discharge of the battery of recorder 
to damage the battery. 

23

22

Содержание HXC-1308

Страница 1: ...Technology Development Zone Qingdao 266510 P R China Web www haiermedical com Version 1st 2016 Dedicated code 0270501029 V13026 Read the Operation Manual carefully before using your appliance Keep the Operation Manual in a safe place Appearance color and layout of the door may vary Translation of the original instruction The scope of the product application For providing a storage environment of 2...

Страница 2: ......

Страница 3: ...rent from the figure in the manual due to product improvement The manual may be subject to changes without further notice Computer panel control digital display of temperature temperature display accuracy as 0 1 C and temperature range as 4 C 1 C temperature range of HXC 1308 and HXC 1308B as 4 C 1 5 c Multi fault alarm high and low temperature alarm low battery alarm power failure alarm sensor fa...

Страница 4: ...fety Precautions Precautions for Use Product Installation Component Names Control Panel Application method Alarm Cleaning and Maintenance FAQ Refrigeration Schematic Circuit Diagram Specification 2 3 6 7 12 15 26 31 33 34 37 Content ...

Страница 5: ...ct to prevent personal injuries and refrigerator damage In case of flammable gas leakage such as coal gas shut off the valve of leaked gas open windows for ventilation and exhaust do not plug in or pull out the power plug of refrigerator to prevent explosion and fire The refrigerator shall only be installed by technician or after sale maintenance staff Otherwise it may cause electric shock or fire...

Страница 6: ...ugged into the socket to prevent fire Apply socket with ground wire to prevent electric shock If the socket fails to be grounded ground wires must be installed by technician Do not use the refrigerator outdoors Electric leakage or shock may be caused if the refrigerator gets wet by rainwater Do not place the refrigerator in humid places or places where the refrigerator may suffer splashing water T...

Страница 7: ...to children And the flammable foaming needs to be disposed by professional persons CP IP foaming materials are flammable need professional processing HXC 608B 6 Drawers HXC 608 Blood Bank Refrigerator External dimensions W X D X H mm Internal dimensions W X D X H mm Effective volume Door Insulating materials Compressor Shelf Drawer Refrigeration mode Outer plate inner plate Condenser evaporator Re...

Страница 8: ...e solvent such as gasoline Shut off the power as well as the battery switch if the refrigerator will be out of service for a long time 39 6 Precautions for Use Specification HXC 608 115V 60Hz Blood Bank Refrigerator External dimensions W X D X H mm Internal dimensions W X D X H mm Effective volume Door Insulating materials Compressor Shelf Drawer Refrigeration mode Outer plate inner plate Condense...

Страница 9: ...W The refrigerator shall be grounded in a reliable manner If the socket is provided with grounding wire check if it is well grounded before operation If the socket is not provided with grounding wire a new one shall be installed by engineering technician Do not ground the refrigerator through gas pipes water supply pipes telephone lines or lightning conductors as electric shock may be caused The p...

Страница 10: ...ailure alarm 48 hours when battery is fully charged I Temperature controller USB function temperature recorder Optional optional Y Electric shock resistance type Power connect style Climate Type Foaming Agent Noise Level 4 42dB A CP IP Note Climate Type 4 means the temperature is 30 the relative humidity is 55 Technical information may vary from the appliance that you have just purchased due to te...

Страница 11: ... control panel Internal fan Display panel Sensor Defrosting Control Lower temperature Upper temperature Alarm VCC GND Heating strip at the cabinet outlet Heating strip of center sill Door heating transformer Door heating strip 1 Temperature recorder LED drive model Lamp switch USB drive panel USB port Reserved network interface Wiring terminal 2 Double throw switch Press wiring terminal box Power ...

Страница 12: ...t the articles gradually into the refrigerator after complete examination of refrigerator 35 10 Circuit Diagram HXC 158 158B 358 358B 608 608B CN2 Cooling fan Compressor Defrosting heating strip CN2 CN1 CN3 CN11 BT1 BAT BAT Main control panel CN5 Internal fan Display panel CN2 CN3 Sensor Alarm Defrosting Control Lower temperature Upper temperature VCC GND Heating strip at the cabinet outlet CN1 t ...

Страница 13: ...erature at one time If so the compressor may not stop for a long time with temperature reducing slowly which will reduce service life Articles must be put in batches and the temperature shall be reduced in steps until it reaches the temperature required Do not damage refrigerating loop Do not apply electronics without product permission in the refrigerator The refrigerator has memory function towa...

Страница 14: ...mean data tested in the standard laboratory without noise in which the refrigerator without storage of articles is laid on the rubber pad and is steadily operated after closing the door to measure out the data at 1 meter from the surface the data do not include the value of startup and shutdown of the refrigerator During the process of usage it is normal that the actual noise is different from the...

Страница 15: ... is received Consult us by phone for detailed information Purchasing of backup battery of the recorder Battery in the same specification and the same brand can be purchased in proper shops 9V DC In order to guarantee the reliability of the performance of the battery you can apply to our company for purchasing The ordering cycle is 1 2 months and it will be delivered 15 days after your money is rec...

Страница 16: ...ipment and may lead to the occurrence of fault Compressor and other mechanical parts are in fully closed state and they do not need to be lubricated Clean the refrigerator Cleaning and Maintenance Component Cleaning In order to prevent from getting an electrical shock or personal injury make sure that the power of refrigerator is cut off before conducting any repair to the refrigerator Make sure t...

Страница 17: ...battery switch is on or off the remote alarm can be started The remote alarm port coordinates with the audible alarm on the refrigerator Thus the remote alarm can be stopped by pressing the key of buzzer outage alarm cannot be canceled COM NO NC Temperature displayer The refrigerator is set to reach appropriate temperature automatically before leaving the factory 4 1 4 1 5 for HXC 1308 1308B Tempe...

Страница 18: ...for the refrigerator He is in charge of examining the running status of the refrigerator and making the record If the temperature inside the refrigerator is too high or too low or other abnormal phenomenon he shall transfer the articles to other refrigerators which conform to the conditions till the fault has been settled and articles can be stored in that refrigerator again When the battery has f...

Страница 19: ...back to the normal display mode The original temperature set value 4 flickers Please do the next operation within 60s Otherwise the display panel will return to display TS SN Alarm items Buzzing Light flashing 3 Fault of sensor Conditions of alarm start Display requirements Note Display board displays the alternation between the mean value of temperature of both upper and lower sensors and E3 wher...

Страница 20: ...after a period of Fd user parameters can be set the default value is 30 min the buzzer and remote alarm will be restarted 1Hz Continuously Buzzing 1Hz Continuously Buzzing Alarm light flashes Display board displays the alternation between the temperature of upper sensor and E2 wherein E2 displaying for 2s while the temperature displaying for 6s HXC 1308 1308B Temperature recorder HXC 158 158B 358 ...

Страница 21: ...ntil the pen returns to the paper At this moment the pen will pause at the most external temperature scaling line of the recording paper this scaling line does not have to be marked with temperature value but must be the one at the most external circle If the pen does not pause at the above position you can use 1 or 2 arrow key to adjust the position within 5 seconds to align the nib to the scalin...

Страница 22: ...ng the change of temperature in the box In order to use the disk recorder in a right and reasonable way please read the instruction bellow carefully Get familiar with your temperature recorder Adjustment of Calibration of the Recorder The recorder is accurately adjusted before leaving the plant It will not be affected even if the power is cut off If adjustment is required please conduct it accordi...

Страница 23: ...nction The computer panel has USB output function which can make use of U disk to export temperature and other test data Plug in U disk the machine begins to automatic identification and import data into U disk During data import press Conform calibration key the flashing display of USB on display panel represents the data export is being on and not yet complete After flashing for 5s it exits and ...

Страница 24: ...r position can also be adjusted finely 3s later the scriber will slide to appointed scale according to the present perceived temperature then the calibration ends After retightening hold down bolt and pressing Key 3 again the recorder will switch to the process of paper calibration Please refer to the explanation of last page to calibrate the installation of paper After calibration the temperature...

Отзывы: