background image

Hager & Werken

 GmbH & Co. KG • Postfach/P.O.B. 10 06 54 • D-47006 Duisburg (Germany)

Tel : +49-(0)203-99 269-0 • Fax: +49-(0)203-29 92 83 • www hagerwerken de

SPEED-MULTIPOL

Light and Hot Air Polymerization Device

For the extra-oral light and heat treatment of crown and bridge material on composite basis, such as ESTENIA

TM

Art.-No. 401 121

Warranty:

The producer ensures that the product is free of defects and that
it shows all guaranteed features and/or all characteristics agreed
upon. The warranty starts with the day of delivery of the product
and is valid for twelve months. Excluded from the warranty are
faults, careless use of the product, connection to unsuitable power
supply, operation with the wrong voltage, acts of God, such as fire,
lightning, humidity, etc. In case of justified complaints the producer
has the right to repair, to deliver spare parts or to decrease the
sales price. Cancellation of the sales  contract can only be claimed
by the customer after two repairs have been done without success.

(

a

)

(

a

)

(

b

)

SPEED-MULTIPOL Technical Data:

Input:

max. 800 W

Operating voltage:

230 V/50 Hz

Fuse:

1 x 8 A

Precision thermostat:

+/- 1

°

 C

Size:

33 x 29 x 47 cm

Weight:

approx. 13.5 kg

Instructions for Use

Mounting:

Take the device out of the packing carefully.
Put the device on a plane surface in such a way that the back of
the ventilation grid stands free, with a distance to the wall of at
least 15 cm.
Lay the fire-clay disc onto the plexiglass plate and fasten it with
the metal ring.
Put the safety plug of the cable into the respective socket. Activate
the main switch to ”position 1”.

Light Chamber (a):

Switch-on the device with the main switch.
Open the light chamber by turning down the lower door.
Put the crown or bridge work to be polymerized onto the plate.
In case you do not want to put in a complete model, but e.g. only
a crown, please use the respective support with fixing pins. Please
make sure the part of the work, that is to be polymerized, points
in upwards position.
Only use support plates designed for the device.
Do not try to raise the work by means of aids.
Shut the light chamber.
Adjust he polymerization cycle according to the quality of material
and polymerization time.
Press the start button.
If necessary, please control if the supporting disc is turning correctly.
The end of the cycle is announced by an acoustic signal.
For ending the signal, press the start-/stop button, open the light
chamber and take out the polymerized work.
Shut the door.
By simultaneous pressing of the min. and sec. buttons, the timer
is re-adjusted to zero.

Heat Chamber (b):

Open the heat chamber by turning down the upper door.
For heat treatment, you have to put the work onto the perforated
supporting plate, eventually in a Petri dish.
Make sure that the work doesn’t get into contact with the chamber
walls!
Shut the heat chamber.
Adjust the desired cycle (temperature and time).
Press the start button.
Open the chamber at the end of the cycle.
Take out the work by means of a pair of tweezers.
Shut the chamber.

Отзывы: