Automatic Operation: Cell Salvage
P/N 53063-50, Manual Revision: B
5-4
The following screen will be displayed:
3. The CS5+ optics and effluent line sensor will monitor the processing of the
cells in the bowl and the quality of the effluent solution. From this
information the CS5+ will automatically determine what the quality of the
incoming blood is and control the processing parameters to optimize the
procedure. In the AUTOMATIC FILL mode, the CS5+ will optimize the
processing by operating at the optimum pump speed between 300 ml/min
and 600 ml/min (or 200 - 300 ml/min if the 125 ml bowl is being used).
Therefore, the pump speed may vary as the bowl is being filled. The CS5+
will operate at default settings when using the 70 ml bowl. Pump speeds
will not vary with the 70 ml bowl.
4. If the reservoir empties before the bowl is full the system reverts to
STANDBY and the display provides this information in the following screen.
Note: The system will resume filling once the preset level has been reached in the
reservoir.
FILL BOWL
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Press MODE twice to enter EMERGENCY option
AUTOMATIC
xx ml
0 ml
0 ml
0
Pump Rate:
600 ml/min
STANDBY
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
to FILL automatically
Press START to FILL bowl or CONC to concentrate
Press WASH to wash partial bowl
AUTOMATIC
695 ml
0 ml
0 ml
0
Need 400 ml in reservoir
Reservoir was emptied
Press RETURN to empty bowl into reservoir