Introduction
P/N 53063-50, Manual Revision: B
1-12
Construction
a.
Cabinet and Cart Materials:
Fabricated sheet metal and cast aluminum.
b.
Control Panel:
Membrane switch technology, environmentally sealed with
backlit option keys.
Note: Only Haemonetics disposable sets may be used with the Cell Saver 5+
System; the system will not function properly with other sets.
Processing Speeds (in ml/min)
Automatic
Emergency
Manual
225 ml bowl
Fill
600
a
800
500
Wash
500
a
800
500
Empty/Return
b
500/300/250
500/300/250
500
Concentrate
450
450
350
125 ml bowl
Fill
300
a
400
500
Wash
300
a
400
500
Empty/Return
b
400/250/200
400/250/200
500
Concentrate
250
450
350
70 ml bowl
Fill
125
125
Wash (from FILL)
100
100
Wash (from CONC)
75
75
Empty/Return
100
100
Concentrate
75
75
Sequester (125 ml or 225 ml bowl)
Draw
60
Empty
b
400/250/200
a.
The machine will adjust the pump speed during the automatic FILL
mode between 300-600 ml/
min (2
00-300 ml/min for a 125 ml bowl) and
during the automatic WASH mode between 250-500 ml/min (200-300
ml/min for a 125 ml bowl) based upon the effluent line sensor readings to
optimize the quality of the final product. Pump speed will not adjust with
the 70 ml bowl.
b.
The machine will adjust the pump speed during the automatic EMPTY/
RETURN mode in preprogrammed increments to assure optimal perfor-
mance.
Содержание Cell Saver 5+