Sequestering using the Cell Saver 5+ System
When the bowl is completely empty, the pump will stop and the following
information will be displayed:
Figure 7-7, Example of the STANDBY state screen display
Transferring the RBCs for reinfusion
The red blood cells can now be drained to a transfer bag for reinfusion to the
patient if needed. The RBCs should be treated the same as a unit of washed,
packed RBCs in terms of administration to the patient and outdate of the
product.
Warning: In accordance with the applicable current guidelines and stan-
dards, it is recommended that a transfusion filter capable of retaining parti-
cles potentially harmful to the recipient be used when returning processed
red cells.
Warning: The operator should refer to the current standards for expiration
date of stored blood.
A PRESSURE CUFF OR ANY OTHER MECHANICAL DEVICE SHOULD
NOT BE USED WITH THE CS5+ SYSTEM. PRESSURE REINFUSION CAN
RESULT IN THE FATAL INFUSION OF AIR INTO THE PATIENT.
If another sequestering pass is desired, the entire procedure may be repeated
following the previous bulleted steps until the desired volumes of PRP and
PPP have been collected.