IEC 60601-1-2, Table 204
Guidance and manufacturer’s declaration―electromagnetic immunity
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
BloodTrack Kiosk V5 is used exceeds the applicable RF compliance level above, the BloodTrack Kiosk V5
should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the BloodTrack Kiosk V5.
b
Over the frequency range 150kHz to 80 MHz, field strengths should be less than 10 V/m.
B.2.4 IEC 60601-1-2, Table 206
The BloodTrack Kiosk V5 is intended for use in the electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
BloodTrack Kiosk V5 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment
(transmitters) and the BloodTrack Kiosk V5 as recommended below, according to
the maximum output power of the communications equipment.
IEC 60601-1-2, Table 206
Recommended separation distances between portable and mobile RF communications equipment
and the BloodTrack Kiosk V5 - for equipment and systems that are not life-supporting
Separation distance according to frequency of transmitter (m)
Rated maximum output
power of transmitter (W)
150 kHz to 80 MHz
d =
P
35
,
0
80 MHz to 800 MHz
d =
P
35
,
0
800 MHz to 2,5 GHz
d =
P
7
,
0
0,01
0,035
0,035
0,07
0,1
0,11
0,11
0,22
1
0,35
0,35
0,7
10
1,1
1,1
2,2
100
3,5
3,5
7
For transmitters rated at a maximum output power not listed above, the recommended separation distanced
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer.
❖
NOTE:
At 80 and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.