BloodTrack Kiosk® V5 Installation Guide
HAE: SM-BloodTrack-08(AA)
SBC100xxx
GBE: 2000-JNB0012
43
7.0
Mounting the Uninterruptable Power
Supply (UPS)
The optional UPS provided is a slimline model, and a custom-built bracket and fittings
are supplied to wall-mount the unit near the BloodTrack
®
Kiosk V5. A suitable patch
mains cordset cable for the UPS mains outlet to the BloodTrack Kiosk V5 mains inlet is
supplied with the BloodTrack Kiosk V5.
Figure 18
UPS mounted below BloodTrack Kiosk V5
Note:
It is recommended that the bracket and fittings securing the UPS be inspected on a
regular basis to ensure the fitment remains secure
—for example, every six months.
The time period for inspection should be determined by suitable risk analysis
undertaken by the responsible authority at the healthcare site.
7.1
UPS Internal Battery
The UPS unit will require maintenance to replace its internal battery after it has been in
service for a period of time. Typically it is recommended the battery is replaced after
three years. This maintenance should be undertaken by a competent person in
accordance with the specific UPS instruction and service manual. Contact
Haemonetics
®
Technical Support if assistance is required
—see
Appendix C